EJA210E flange mounted differential pressure transmitterAdopting monocrystalline silicon resonant sensor technology, suitable for measuring the density of crystalline solids or precipitated liquids. EJA210E converts the measured differential pressure into a 4-20mA DC current signal output, which can measure, display, or remotely monitor static pressure. It has functions such as fast response, remote setting, and self diagnosis.
The EJA-E series products offer BRAIN, HART/HART (1-5VDC) low-power type, FF fieldbus, and PROFIBUS PA communication protocols, with standard configuration certified with SIL 2 safety.
EJA210E flange mounted differential pressure transmitter output:
4-20mA HART/BRAN (output signal code D&J); Two wire system with digital communication, 4-20mA DC output, can be set as linear or square root. BRAN or HART
The FSK protocol is loaded on a 4-20mA signal, with an output range of 3.6mA to 21.6mA. To comply with NAMUR NE43, output limits can be pre-set through option codes C2 or C3.
1-5V HART (output signal code Q); three wire or four wire system 1-5V low-power type can be set to linear or square root. The HART protocol is loaded on 1-5 V signals, with an output range of 0.9-5.4VDC
power supply voltage
4-20mA HART/BRAN (output signal code D&J)
10.5~42V DC ordinary type and explosion-proof type
10.5-32V DC with lightning protection (option code/A)
10.5~30V DC intrinsic safety, n-type, non flammable type
● Digital communication (BRAIN and HART): * Small 16.6V DC; 1-5V HAR Ding (output signal code Q)
9-28V DC ordinary type and explosion-proof type
Shenyuan consumption: 0.96-3mA, 27mW
Environmental temperature: -40~85 ℃ (-40~185 ° F); -30~80 ℃ (-22~176 ° F) with LCD meter head
Process temperature: -40~120 ℃ (-40~248 ° F)
Protection level: IP66/IP67, NEMA 4
Flange dimensions: 2-inch (50mm), 3-inch (80mm), 1 1/2-inch (40mm), 4-inch (100mm)
EJA210E flange mounted differential pressure transmitterSelection: