Applied in laboratories or soil for measurementO₂concentration

Typical applications:
lIn the measurement laboratory、Gradient soil or undergroundsoilOxygen in the middle;
lMonitoring oxygen in indoor environment for climate control, used in composting and tailings;
lMonitoring the oxidation-reduction potential in soil;
lCalculate the consumption of oxygen in a closed environment by measuring the respiration rate of oxygen.
Product features:
ØSignal output (optional)
0---50mV tributary voltage
SDI12
Handheld reading meter
ØHeating detector
When the sensor is used in soil or composting environments, the heated breathable membrane at the inlet can not only prevent the condensation of water vapor, but also block the diffusion path of water vapor.
ØUnique design
There is a fully sealed thermistor inside the sensor to monitor temperature. The sensor is placed inside the polypropylene valve body and can be used underground for a long time. Two optional heads include a diffusion head for outdoor environments and a flow-through head for laboratory environments.
ØSimple calibration
The output voltage is proportional to oxygen. Calibration is achieved by measuring the voltage under standard environmental conditions(O ₂ atmosphere is 20.95%) and derive a linear correction factor. Zero offset N ₂ gas can be obtained (recommended measurement below 10% O ₂).
Technical parameters:
model: |
SO-100 |
SO-200 |
characteristic: |
Diffuse type, can be used for soil |
Circulation type, can be used for rapid measurement |
Size: |
3.5 cm in length and 3.5 cm in diameter; 125 mesh aperture |
3.2 cm in length and 3.2 cm in diameter; Pipeline interface: 1/8 " |
Range: |
0-100% O2 |
|
Accuracy: |
Daily Drift<> |
|
Repeatability: |
±0.001% O2 (10 ppm) |
|
Input power supply: |
12V power supply, used for heaters; 2.5V excitation is used for electric heating regulators |
|
Output: |
50 mV at 20.95% O2 |
12 mV at 20.95% O2 |
life: |
Continuous use for more than 10 years |
Continuous use for more than 5 years |
Response time: |
60 seconds |
12 seconds |
Operating environment: |
Temperature:-20-60 °C,Humidity: 0-100% RH, air pressure: 60-140 kPa |
|
Weight: |
210g |
|
Cable: |
2m cable |
|
Gas impact: |
No impact: CO2, CO, NO, NO2, H2S, H2,CH4; Impact<1%: NH3, HCI, C6H6 (benzene), the sensor is very sensitive to SO2, and in O3, the sensor may be damaged. |
|
Manufacturer: Apogee, USA