Fully automatic multi parameter chemical analyzer
Smartchem Discrete Auto Analyzer

SmartchemFully automatic chemical analyzer, also known as discrete chemical analyzer or intermittent chemical analyzer. It is a new breakthrough in the field of ion analysis, with internationally advanced technology. The entire measurement process is controlled by a computer, with a high degree of automation and simple operation. The detection and analysis results are accurate and reliable, in line withISO、AOAC、BATF、USEPA、AFNOR、COFRACWidely used in water quality monitoring according to international certification standards(Including seawater, domestic sewage, industrial wastewater, groundwater, river water, drinking water, etc)In the fields of food analysis, alcohol analysis, soil solution analysis, etc., it is currently the world's new generation of ultra advanced fully automatic chemical analyzers.
working principle:
Use the classic colorimetric method. Using robot technology, the fully automatic sampling needle adds reagents and samples into the colorimetric cup, stirs and mixes well, and reacts. Then use a high-precision dual beam digital detector to measure absorbance. For different measurement parameters, Smartchem can automatically select the corresponding wavelength and measure without the need to replace modules. Conventional measurement customers only need to arrange the test sequence, load reagents and samples, and then the machine enters automatic measurement mode.
Product features:
Simultaneous measurement of multiple parameters:Smartchem can measure multiple parameters simultaneously. Traditional flow analysis techniques require additional channels for measuring different parameters, with several channels needed for each parameter or different modules replaced. andSmartchemBreaking channel restrictions, only one host is needed to solve multi parameter measurement problems, saving costs. contain9An optical filter and1A blind spot filter. Spectral range:340nm~880nmOne injection can immediately obtain all measured parameters.
Low consumption, cost savings, and reduced pollution:The computer-controlled micro pipette controls liquid dosage (microliter level), with fast measurement speed, lower consumption of reagents and samples, less harmful waste liquid, cost savings, and environmental protection.
The sample has no cross contamination and achieves high precision:SmartchemAll kinds of samples and reagents are stored and taken independently. All samples in traditional flow analyzers enter through the same channel and are isolated only by water columns or bubbles, which inevitably leads to cross contamination between samples and affects measurement accuracy. andSmarchemDue to the use of sampling needles for sampling, the above-mentioned defects are completely avoided. After each liquid sampling, the sampling needles are automatically and strictly cleaned to prevent cross contamination and carry over contamination, thus improving the accuracy of measurement.
Nitrate salt-Nitrite specific module(Optional):For nitrates-Analysis of Nitrite, Applied in the United StatesEPAAuthoritative analysis method-Cadmium column reduction method(USEPA 353.2),The reciprocating peristaltic pump promotes turbulence, fully reacts, and maximizes the contact between nitrate ions and cadmium. The cadmium module can be replaced on site at any time.
Fast measurement speed:Due to the use of microanalysis methods,SmartChemAchieved exceptionally short analysis time, significantly improving the sample analysis speed of wet chemical analysis methods. During the measurement process, the sample holder can be replaced without interrupting the analysis, achieving continuous measurement.
Advanced photometer:Adopting high-precision, dual beam digital detector, the extended linear range of absorbance values reaches2.500AUFSReduce the reanalysis of sample overages. Multiple detection modes:endpoint, fixed time, differential, bichromatic, and kinetic(Endpoint detection, timed detection, differential detection, dichromate detection, kinetic detection).
Easy to operate, graphical display, clear at a glance:By using visual operations, the simulation instrument interface settings parameters are clearly visible, reducing the possibility of human errors.
Fully automated measurement:Fully automated, automatic method switching, automatic standard preparation, automatic dilution, fully integrated, no need to choose a diluter. The colorimetric cup is automatically cleaned and can provide continuous measurement. Before each use, a colorimetric cup test is conducted to ensure cleanliness. If the measurement fails, the unclean colorimetric cup can be skipped. Easy to operate, standard sample configuration, automatic conversion between methods, and automatic dilution of over range samples are all automatically completed by the instrument.
Multiple measurement methods can be freely added:The instrument can have hundreds of measurement methods built-in, and can also be programmed and set by oneself. Capable of performing wet chemical analysis methods similar to manual or SFA/FIA analysis techniques, compatible with methods such as USEPA, ASTM, AOAC, DIN, and ISO.
Application areas and testing projects:
Water quality monitoring: acidity, alkalinity, ammonia nitrogen, chemical oxygen demand(COD)Chloride, chlorine, chromium, cyanide, fluoride, hardness, nitrate, nitrite, volatile phenol, phosphate, silicate, sulfate, total Kjeldahl nitrogen, total Kjeldahl phosphorus, aluminum, iron, manganese, urea
soil&Plant analysis: ammonia nitrogen, nitrate, nitrite, phosphate, total nitrogen, total phosphorus, chloride, boron, calcium, magnesium, lysine, urea
feed&Fertilizer analysis: ammonia nitrogen or protein, calcium, phosphate, total nitrogen, total phosphorus, urea
Tobacco analysis: ammonia nitrogen, chloride, nicotine, nitrate, phosphate, cyanide and formaldehyde in tobacco
Alcohol analysis: glucose, fructose, sucrose, malic acidtartric acidAcetic acid, Portugal(Grape wine)Sugar acid, iron, copper, total aciditypHTotal sulfur dioxide, free sulfur dioxidePANOPAAmmonia nitrogen
For other unspecified application areas, please consult our technical personnel.
Technical indicators:
1Optical detector:
Spectral range:420nm-880nm,
Dual beam digital detector:
Light source:6V/10WHalogen lamp
Optical path length:10nm
Linear range:0.001-2.500Abs
resolving power:0.0001 Abs
Expand the absorption range to2.500AUFSReduce the reanalysis of sample excess range.
2Filter wheel:9An optical filter and1A black blind spot filter
3Test input mode:Optional mode, graphic mode, batch processing mode, emergency processing mode
4Reagent and sample grade, injection system:
2Up to one liquid storage rack16A reagent position, among which10A refrigerated reagent position,6A room temperature reagent position,10A standard sample position and2A diluent position
5 sample racks, suitable for continuous placement of samples in batches, can be placed at once126A sample
1A robotic arm provides all injection operations: the computer can control a high-precision micropipette with a pipetting volume of3 - 450 μLAfter each sampling, it can be automatically cleaned and dried, and the same scale sensing sampling needle can maintain
The sample size is controlled within:3.0ul-300ul
The amount of reagents should be controlled at:3.0ul-300ul
5Dilution device: does not use traditional syringes for dilution, dilution can reach100times
6Reaction disk:60A semi reusable colorimetric cup and plate with automatic cup cleaning and continuous computer-controlled cup quality
Heating temperature:
7Nitrate salt-Nitrite module:SmartChemThe instrument method fully followsUS EPAmethod#353.2
8Operating environment:
Electricity requirements:90-250VAC,47/63Hz,300W
Water consumption:2ml/test
Environmental temperature:
Relative humidity:20%-85%
9. DivideAnalysis module:
High:
Width:
Long:
Heavy:
10Software
WindowsOperating system compatibility, powerful analysis software
There are pre-set analysis methods in the software
Automatic dilution settings for samples and standards
Dilute the sample beyond the measuring range again and measure it
The standard sample can be manually prepared or an original standard sample can be provided for automatic preparation
Create a quality control chart through quality control samples
Manufacturer and place of origin:AMSgroup/Italy