The German Ratfisch online methane total hydrocarbon analyzer RS200 is suitable for emission monitoring, online measurement of total hydrocarbons/methane, and process optimization. The typical features of this laboratory include a heated analysis chamber (adjustable to 200 ℃), automatic conversion of methane and total hydrocarbons, and automatic and range setting of means.

Product Details

Application areas of the German Ratfsich online methane total carbon hydrogen analyzer RS200

Emission Monitoring

Online measurement of total hydrocarbons/methane

process optimization

typical characteristics

Heated analysis room (adjustable up to 200 ℃)

Automatic conversion of methane to total hydrocarbons

Automatic and means range setting

Flame extinguishing indicator

Automatically turn off fuel

Automatic monitoring capability (temperature, pressure, flame)

Optional sampling pipeline

Optional filters

Optional built-in zero point gas and combustion air supply device

Optional data collection software

Technical parameters of the German Ratfisch online methane total carbon hydrogen analyzer RS200

Range: 0~10/100/1000ppm

Linear: better than ± 2% of the full range deviation

Cross sensitivity:<± 4% of full range deviation

Detection limit:<2% of full range deviation

Zero drift:<2% of full range deviation per week

Span drift:<4% of full range deviation per week

Response time: T90 approximately 5 seconds (without sampling pipeline)

Sampling flow rate: heated sampling air pump, approximately 0.5 liters per minute

Filter: 4 μ/100cm2 stainless steel sample gas filter

Analysis temperature: adjustable from+45 ℃ to+200 ℃

Pipeline temperature: adjustable from+45 ℃ to+200 ℃

Electromagnetic valve: Zero gas and standard gas electromagnetic valve

monitor:

Measurement value: 5 and a half LED (resolution of 3 and a half bits)

Parameter: 3-digit LED

Status: LED indicator bar

Fuel consumption: 0.9 liters per hour (H2 fuel @ 200 ℃ analysis temperature)

Combustion air: 10 liters/hour (analysis temperature of 200 ℃)

Catalytic conversion: temperature controlled at+450 ℃

Signal output: 0~10V or 0/4~20mA

Status touchpoint: Pre alarm (option)

Main alarm (option)

Range recognition

Flame extinguishing indicator

State recognition (zero gas, standard gas calibration)

Alarm (pressure, temperature, flame)

Environmental temperature:+5 ℃~+35 ℃

Power supply: 230 V/50 Hz@800 tile

Dimensions: Width 19 "x Height 267mm x Depth 500mm