MMD-HR120 roughness profiler

NegotiableUpdate on 01/30
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
MMD-HR120 roughness profiler X stroke: 120mmZ1 range: 50mm X axis: (1+L/300) mZ1 axis: (0.8+2H/100) m
Product Details
Measurement project
Roughness section Ra,Rz=Ry(JIS),Rq,Rt=Rmax,Rp,Rv,R3z,R3y,RzJIS,Rs,Rsk,Rku,Rsm,Rmr
Contour meter section Point, line, circle, partial circle, ellipse, arc processing (arc radius, center distance, distance from center to line, distance from intersection to center, etc.), highest point, lowest point, distance, angle processing (coordinate angle, line angle), straightness, groove depth, groove width, groove edge distance, groove center distance, horizontal distance, convexity (logarithmic curve), contour degree and other parameters
Roughnessmeasuring range
x-axis RA:0.005μm—16.00μm
Z1 axis ± 20 μ m, ± 40 μ m, ± 80 μ m
Sampling length L 0.25mm, 0.80mm, 2.50mm, automatic
Evaluate length 1L-5L (optional L is the sampling length)
Measure travel length 3L -7L (optional L is the sampling length)
Contouring instrumentmeasuring range
X-axis (horizontal guide rail) 120mm
Z1 axis (sensor) 50mm
Z-axis (column) 400mm
Roughnessmeasurement accuracy
Indication error 1 0.001μm
Indication error 2 ≤±7%
Contouring instrumentmeasurement accuracy
x-axis
(L=X-axis guide rail movement distance)
±(1+L/300)μm
Z1 axis
(H=Z1 axis measurement height)
±(0.8+2H/100)μm
XAxis horizontal arm
Straightness 1μm/120mm
Grating resolution 0.1μm
Origin of grating Import
Z1Axis sensor
resolution 0.05μm
sensor type Grating sensor
origin Import
Movement direction Front/rear dual direction
dynamometry 0.5-6g
Measuring needle tip Hard alloy R: 0.05mm, 0.01mm
Climbing angle Rising 77 degrees and falling 88 degrees
Drive travel
Maximum driving stroke length 17.5mm/0.71 inch
Minimum driving stroke length 1.3mm/0.052 inch
Roughnessspeed
measurement speed 0.135mm/s, 0.5mm/s, 1mm/s
Contouring instrumentspeed
measurement speed 0.1mm/s, 0.2mm/s, 0.5mm/s (custom)
Positioning speed 0.1mm/s, 0.2mm/s, 0.5mm/s (custom)
Magnify graphics/reduce
zoom Can be arbitrarily reduced or enlarged
Comprehensive adjustment platform
Comprehensive adjustment platform X-axis horizontal tilt adjustment range: ± 30 °
Y-axis direction adjustment range: ± 7.5mm
Z-axis rotation adjustable 360 degrees (with fine-tuning device)
INSTALLATION REQUIREMENTS
power supply 220-240V AC, 50HZ, separate ground wire; Maximum power consumption 450W
compressed air Pure compressed air: pressure 0.4MP, consumption 0.1m3/min (air supply 0.6MP, 0.25m3/min or above)
Environmental requirements Working temperature: 10 ° -30 ° C (normal temperature); Humidity 20-80% RH (no condensation), good grounding, no obvious vibration source