KY-3000SUV fluorescence sulfur analyzer
KY-3000SUV fluorescence sulfur analyzerIt is a sulfur analyzer widely used to determine the total sulfur content in products such as finished oil, crude oil, distillate oil, petroleum gas, plastics, and petrochemical products. Compared with similar instruments at home and abroad, it has the characteristics of wide linear range, low detection limit, strong anti-interference ability, easy operation, stable and reliable results, high analysis accuracy, and good repeatability. It is currently one of the more advanced total sulfur analysis instruments in China. This instrument is equipped with a dedicated quartz tube(number: ZL 2010 2 0242233.X)Specialized cracking electric furnace(number: ZL 2010 2 0242234.4)Obtained utility model certificate.
Main features of the instrument:
vThe instrument adopts the internationally popular USB port communication, standard 24 bit data acquisition, and the minimum detection signal is less than 1 microvolt.
vData can be stored, accessed, printed at will, and relevant problems can be solved based on on-site data files.
vThe main components of the instrument are made of original imported components, which have better overall performance than similar domestic instruments and can replace imported products.
vThe high voltage of the photomultiplier tube (PMT) can be adjusted arbitrarily, and the calibration of the standard curve can be done using single point or multi-point calibration, making it easy to operate.
vThe software can run on all current operating systems and has strong adaptability. The instrument software has powerful functions.
Instrument execution standard:
vSH/T 0689, ASTM D5453 Determination of Total Sulfur Content in Light Hydrocarbons, Engine Fuels, and Other Oil Products (Ultraviolet Fluorescence Method)
GB/T 34100-2017 Determination of Total Sulfur Content in Light Hydrocarbons, Engine Fuels, and Other Oil Products by UV Fluorescence Method
Main technical parameters:
vSample types: liquid, solid, and gas (with corresponding injectors required)
vMeasurement method: UV fluorescence method (S)
vMeasurement range: 0.2mg/L~5000ppm~percentage content
vDetection limit: 0.2mg/L (sulfur element)
vRepetitive error: ① 0.2mg/L≤X<1.0, ≤ ±0.1mg/L;
② 1.0mg/L≤X<100mg/L, ≤ 10%;
③ 100mg/L≤X≤10000mg/L, ≤ 5%
vGas source requirements: High purity oxygen: above 99.98% High purity argon: above 99.98%
vTemperature control range: room temperature~1300℃
Instrument system configuration:
vNormal configuration: KY-3000S UV fluorescence sulfur analyzer complete set (including computer, quartz tube, liquid sampler, etc.);
vImported accessories: photomultiplier tube (Japan), data acquisition card (USA), membrane dryer (USA), filter (USA), UV lamp (USA), etc;
vOptional accessories: solid sampler, gas sampler, quartz tube cracking, and various standard samples.