KY-102 Infrared Sulfur Carbon Analyzer

NegotiableUpdate on 05/08
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

KY-102 Infrared Sulfur Carbon Analyzer: It is a new type of carbon content determination instrument based on the principle of infrared detection. After high-temperature combustion, the carbon in the sample is converted into CO2, and the sulfur is converted into SO2. CO2 and SO2 are introduced into the infrared detection cell. Due to the absorption of a certain amount of infrared energy by CO2 and SO2, the signal at the instrument detection end changes. According to Lambert Beer's law, the carbon and sulfur content in the sample can be calculated.

Product Details

Introduction to KY-102 Infrared Carbon and Sulfur Analyzer:
KY-102Infrared carbon content analyzerIt is a new type of carbon content analyzer based on infrared detection principle. After high-temperature combustion, the carbon in the sample is converted into carbon dioxide, and the sulfur is converted into sulfur dioxide. Carbon dioxide and sulfur dioxide are introduced into the infrared detection cell. Due to the absorption of a certain amount of infrared energy by carbon dioxide and sulfur dioxide, the signal at the instrument detection end changes. According to Lambert Beer's law, the carbon and sulfur content in the sample can be calculated. Compared to traditional chemical methods, instruments have the advantages of being fast, easy to prepare, and do not require any chemical reagents. They are used for the determination of carbon content in catalysts.

Infrared carbon sulfur analyzercharacteristic:

vAnalysis method: Infrared absorption method

vSample weight:100mg1g

vRequired gas: Oxygen (pressure)0.5Mpa,purity99.7%

vSingle analysis time: approximately1minute

vSample combustion method: viscous sample (tube furnace method)

Solid sample (electric arc furnace method)

vRepetitive error:The concentration is:200ppm1000ppmSample, relative error: ±200ppm

The concentration is:1000ppm5%Sample, relative error: ±10%

vDetection range:200ppm5%