KY-100 (5) Standard COD Dissolver (5 Cone Bottles)

NegotiableUpdate on 05/08
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

The KY-100 standard COD digester uses microelectronics technology for timed control of the heating electric furnace. It can replace the ball reflux tube with a burr reflux tube for 5 or 6 conical bottles (8, 10, or 12 column bottles), and replace the tap water cooling method with air cooling technology. This not only saves water but also standardizes the instrument. They can maintain the necessary high temperature for the reaction within the set time to ensure complete digestion. It can be widely applied in fields such as environmental protection, scientific research monitoring, and production monitoring.

Product Details

KY-100 standard COD digesterBy using microelectronics technology for timed control of heating electric furnaces, it is possible to5 or 6A conical bottle8. 10 or 12 column bottle)The burr reflux tube replaces the ball reflux tube, and the air cooling technology replaces the tap water cooling method, which can save water and standardize the instrument. They can maintain the necessary high temperature for the reaction within the set time to ensure complete digestion. It can be widely applied in fields such as environmental protection, scientific research monitoring, and production monitoring.

Main features of the instrument:

  • Following the basic principles of international standards (ISO) and national standards (GB), the digestion operation with reflux heating and micro boiling for 2 hours is ensured
  • The preparation and dosage of the reagent solution are consistent with the GB method * to ensure reliable and accurate analysis results
  • Simple, accurate analysis results, water conservation, automatic shutdown after completion

Instrument execution standard:

  • HJ828-2017 Determination of Chemical Oxygen Demand for Water Quality - Dichromate Method (Alternative to GB11914-89)

Main technical parameters:

  • Sample type: Liquid
  • Measurement range: 0-1000mg/L, 0-10000mg/L (diluted water sample)
  • Measurement time: ≤ 2h
  • Measurement error: The relative standard deviation of potassium hydrogen phthalate standard solution (500mg/L) is not more than 5.0% for industrial use

The relative standard deviation of organic wastewater (500mg/L) is not more than 8.0%

  • Environmental temperature: 5-40 ℃
  • Heating temperature: adjustable from 25 ℃ to 170 ℃ (± 1 ℃)
  • Heating time: can be set from 0 to 180 minutes
  • Heating power: 1500W (AC 220 ± 22V, 50HZ)

Instrument system configuration:

  • Normal configuration: KY-100 standard COD (cr) digester complete set and glass burr reflux tube;
  • Optional accessory: Dissolving tube.