E-mail
1095969861@qq.com
-
Phone
1392422246313318771005
-
Address
3rd Floor, Building 1, Zhanqing Industrial Park, Haiyong Road, Panyu District, Guangzhou City
Guangzhou Quanfeng Environmental Protection Equipment Co., Ltd
1095969861@qq.com
1392422246313318771005
3rd Floor, Building 1, Zhanqing Industrial Park, Haiyong Road, Panyu District, Guangzhou City
Application:
Widely used in RO, pure water, ultrapure water, circulating water, boiler water and other systems
And in the process fields of electronics, electroplating, chemistry, food, pharmaceuticals, etc
characteristic:
144X144 microcomputer designed conductivity/resistivity transmitter
Outdoor waterproof design suitable for wall mounted, panel mounted, and circular pipe installation
Large LCD display screen with operation status and special symbol display
Equipped with backlight sensor, automatic and manual backlight setting function
■ Provide electrode automatic cleaning device contacts, program setting output
Equipped with automatic and manual range switching, accepting bipolar and quadrupole electrode inputs
■ Manual/automatic selection of temperature compensation function
Password lock and combination key enter the setting and calibration mode design, enhancing security protection
Equipped with RS-485 interface output and Modbus standard communication protocol (EC-4300RS only)
■ Equipped with two 0/4~20mA outputs corresponding to Cond/Resi and Temp settings. (EC-4300 only)
Specifications:
■ Measurement range
Cond: 0.000μs/cm~199.9ms/cm
Resi: 0.00~19.99 MΩNaN
Sal.: 0.0~70.0ppt
Temp.: –30.0~+130.0℃
■ Resolution
Cond: 0.001μs/cm
Resi: 0.01 MΩNaN
Sal.: 0.1ppt
Temp.: 0.1℃
■ Instrument accuracy
Cond: ±1%±1 Digit
Resi: ±1%±1 Digit
Sal.: ±1%±1Digit
Temp.: 0.2℃±1Digit
Temperature compensation: Linear and nonlinear selection at 25 ℃
Automatic temperature compensation Pt1000, NTC30K dual-mode selection
■ Manual temperature compensation button setting
Electrode constants 0.01, 0.05, 0.1, 0.5, 250px -1 fixed
0.001~19.99 cm -1 can be set arbitrarily
Working temperature: 0-50 ℃
Storage temperature -20~70 ℃
■ Large LCD display screen with dual indicators of main parameters and temperature
Analog output with two isolated 0/4-20mA outputs
Program settings corresponding to main parameters/temperature
(EC-4300 only)
Digital output RS-485 Modbus standard communication protocol
(EC-4300RS only)
Relay output high and low points, hysteresis can be set by any program
Two sets of ON/OFF outputs
240VAC maximum current 0.5A
■ Cleaning contact: ON 0~9999 seconds/OFF 0~999.9 hours
A set of ON/OFF outputs
240VAC maximum current 0.5A
■ Power input 100-240VAC wide power supply, 50/60Hz
Installation methods include wall installation, panel installation, and circular pipe hanging installation
■ Instrument size 144mm × 144mm × 115mm (H × W × D)
■ Hole size 138mm × 138mm (H × W)
■ Protection level IP65
■ Weight 0.8Kg