YSI has launchedProODO、ProOBODThe use of non oxygen consuming fluorescence detection technology can provide more stable, repeatable, and sensitive dissolved oxygen detection data, effectively reducing sensor drift.
Widely used in routine and emergency measurement of water environment, dissolved oxygen measurement in sewage treatment, teaching and research in universities and research institutions, water quality measurement in aquariums and aquaculture, ecological environment investigation, and dissolved oxygen measurement in other fields.
Product Features:
Measurement range(0-500%Compared to other brands of optical dissolved oxygen sensors, it has a wider range
Intelligent digital sensor technology can store calibration data in the sensor, even if the probe is installed on any uncalibrated deviceProODOOn the host, it can also run normally
adaptation300milliliterBODbottle
Users can replace sensors themselves in the field
IP67Waterproof level, the battery compartment and instrument circuit compartment are independently separated and sealed, so even if the battery compartment is flooded, it will not affect or damage the instrument circuit
The anti slip casting shell coated with rubber makes it easier to grip and has a longer service life
storage2000Group data (including sensor data, date, time, site, and user-defined information)
With the help of standard equipmentUSBwire connectionProPlusThe host and computer can not only provide power to the host, but also use easy-to-useData ManagerSoftware settings host, management analysis data, and viewing graphical and tabular data
Two measurement modes: automatic/Manual sampling mode/Continuous real-time measurement mode
Luminous keyboard and backlight display screen facilitate operation in low light environments
Users can choose their own cable length, with a maximum length of up to100Rice; Cable sorting kit helps organize4Meter or longer cables
MSMilitary connector, quick plug and play, waterproof, reliable and stable connection
The cable part of the joint can withstand30Ten thousand bends, durable and long-lasting
Built in barometer
Advantages of optical dissolved oxygen sensors:
Unrestricted by flow rate, no flow rate or stirring dependence
No preheating time required
Low drift and long-term stability, no need for frequent calibration
Not affected by hydrogen sulfide or other gases
Lower maintenance: No need to clean electrodes or replace solutions, only need to replace sensitive components once a year
Technical Specifications:
parameter |
measurement range |
resolution |
accuracy |
%Air saturation |
0to500% |
0.1% |
0to200%: Reading ±1%or1%Air saturation, whichever is higher;200to500%: Reading ±1.5% |
milligram/rise |
0to50milligram/rise |
0.01milligram/rise |
0to20milligram/Upgrade:0.1milligram/± of the rise or reading1%Whichever is larger;20to50milligram/Rising: ± of reading1.5% |
temperature |
-5to70℃ |
0.1℃ |
Accuracy ±0.2℃ |
atmospheric pressure |
50-110kilopascal |
0.01kilopascal |
±0.2Kilopascals(0-50Degree) |
Other technical specifications:
working conditions |
-20to70degree |
interface |
With a lock buckleMSMilitary standard waterproof interface |
Salinity input range |
0to70ppt(Manual input value with automatic compensation) |
data memory |
2000Group data (including sensor data, date, time, site, and user-defined information) |
reaction time |
T90at25In seconds,T95at45second |
calibration |
Dissolved oxygen: a little or a little0%Two point calibration; Air pressure: One point calibration |
Certified |
RoHS、CE、WEEE、C-Tick、VCCI、IP67and1 Rice Drop Test |
communication connection |
USB2.0, including communication base andUSBline |
data management |
Data ManagerChinese software, customizable100 A folder and site name |
user password |
To ensure data security, users can set their own password |
GLPfunction |
Save calibration records that comply withISO9000Requirements for certification laboratories |
power supply |
4section2Alkaline battery(25℃, exceeding2000Hour) |
cable |
Provided1、4、10、20、30、40、50、60、100Rice is available for selection |
external dimensions |
21.6×8.3×2.3(Length, width, and height,cm) |
weight |
0.48Kilogram (including battery) |
standard configuration |
communication protocolcommunications saddle(605604)TheUSBConnection cable, software(605515) |