XF-3241 Aircraft Fuel Dynamic Thermal Oxidation Stability Tester

NegotiableUpdate on 01/09
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

XF-3241 Aircraft Fuel Dynamic Thermal Oxidation Stability Tester

Product Details

XF-3241 Aircraft Fuel Dynamic Thermal Oxidation Stability Tester
XF-3241 Jet Fuel Dynamic Thermal Oxidation Stability Tester

XF-3241

ASTM D3241

The XF-3241 Jet Fuel Dynamic Thermal Oxidation Stability Tester is a new analyzer suitable for measuring the dynamic thermal oxidation stability of jet fuel. This instrument has the advantages of small size, light weight, small dosage, short test cycle, and high analysis accuracy. This method uses a 600ml sample to flow at a flow rate of 3ML/min through the surface of a heated test tube at 200-370 ℃, pass through a test filter, and return to the upper part of the oil cylinder. The filter is a porous precision oil filter that can capture the products produced by the deterioration of fuel oil during testing. The degree of deposition of deteriorated materials is represented by the magnitude of the pressure difference before and after the filter, and the magnitude of the pressure difference and the color grade of the surface sediment on the test tube are used as the criteria for evaluating the dynamic thermal oxidation stability of jet fuel.
The working and analysis principles of the instrument meet the requirements of ASTM D3241 and the standard method of GB/T 9169 "Determination of Thermal Oxidation Stability of Jet Fuel (JFTOT Method)", and are carried out according to the method. The measuring instrument can be divided into two main parts; Analyze the system and control system.


technical specifications
Nitrogen pressure regulation range: 0KPa -3.45MPa (6kg/cm2)
Temperature control range: 20-380 ℃
Constant temperature accuracy: 20-380 ℃<± 1 ℃
Nitrogen pressure regulation range: 0KPa -3.45MPa (6kg/cm2)
Fuel system pressure: 3.45 ± 0.345MPa
Differential pressure measurement accuracy: 0-340mmHg, close to level 0.3
Differential pressure display resolution: 0.1mmHg
Test sample size: 600mL
Fuel oil flow rate: 20 drops/9 ± 1 second, 3.0 ± 0.3mL/min
Test time: 150min ± 2 minutes
Cooling water: Flow rate: 38 ± 8L/H Cooling water pressure: 0.1-0.4Mpa
Environmental temperature: 10-40 ℃
Relative humidity of the air:<85%
Power supply voltage: AC 220V ± 10%
Thermostatic bath size: 500 × 190 × 424mm (length × width × depth)
Dimensions: 550 × 450 × 600mm (length × width × height)

Instrument features:
1. Large screen full touch LCD display, which can intuitively operate and display the entire process of flash point test. The interface is simple, and you can directly click to set the experimental mode. It is convenient and reliable to use, and the experimental process and measurement data are clear at a glance.
2. The main electrical components of the control and measurement departments are all imported, ensuring the reliability and stability of the instrument. The instrument has a compact and small structure, and a beautiful appearance.
3. The instrument has self fault diagnosis and alarm functions.
4. The instrument is equipped with a standard needle printer, which can list the test results one by one.