Where to repair the malfunction of the digital display part of Kunshan Sanfeng dial indicator and micrometer

NegotiableUpdate on 09/25
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

The main reasons for the misalignment of the dial gauge are normal wear and damage caused by improper use and storage. Normal wear and tear can be adjusted through the adjustment mechanism inside the dial gauge without disassembly. However, if the dial gauge is damaged or stuck due to improper use or collision, it needs to be disassembled. It is very important to master the correct disassembly method because the parts of the ROK dial gauge are relatively accurate. Disassembly is divided into three steps: * step disassembles the reading device, the second step disassembles the transmission device, and the third step disassembles the body part. Assemble in the reverse order of the above. The main contents of micrometer repair include appearance, interaction between various parts, stability of indication, and indication error

Product Details

Suzhou Lefeng Precision Instrument Co., Ltd(German Japanese Maintenance Department)We are a professional service center specializing in measuring tools and instruments for brands such as Sanfeng in Japan and Mal in Germany. The company has strong technical strength, with multiple engineers from the world's largest measuring tool manufacturing enterprises. The accessories are imported with original packaging, and the measuring laboratory instruments are all imported measuring instruments to solve the problem of poor accuracy after maintenance for customers. The quality service is supported by the metrological quality inspection department, which issues relevant maintenance product inspection reports and cooperates with customersGeneral agent of "Mitutoyo" and "Mahr"Establish long-term stable cooperative relationships. It is one of the most professional measuring instrument maintenance service providers in mainland China.
karmarepairCalipers, height gauges, micrometersdial indicatorMicrometer, internal and external diameter gauge, thickness gauge, roughness meter, 2D height gauge, tool microscope, etc.
Service commitment: Provide free on-site pickup, free fault detection, and within 2 days of quick repair service
Quality assurance: The accuracy of the repaired products meets the national precision standards
Warranty period: Non human damage, 3-month warranty for similar faults

Some common faults and troubleshooting methods of ROK digital dial gauge:

1、 Cleaning
The majority of the workload in the periodic calibration of the ROK dial gauge in use is cleaning and oiling. Generally, remove the back cover and reading device, and place the watch body and movement in 120 # gasoline for cleaning
Wash. When cleaning, push the measuring rod while shaking the dial gauge. After cleaning, blow dry and apply oil to increase lubrication and reduce noise.
2、 Adjustment and repair
1. The axial movement or radial swing of the bezel and body can be adjusted by bending the leaf spring and bow spring slightly towards the back cover, which is equivalent to increasing the thickness of the spring and reducing the axial movement of the bezel. Extend more points of the leaf spring to increase the curvature of the bow spring, widen the spring device radially, and reduce the radial swing of the dial. For other types of micrometers, although the methods may not be the same, the principle is the same, which is to increase the thickness and width of the spring device to reduce the axial and radial movement of the dial.
2. The movement of the measuring rod is not flexible: mainly considering the following aspects
(1) Liquid enters the interior and the measuring rod feels sticky. Clean it with gasoline. Pull the measuring rod out of the meter body and clean the inner walls of the upper and lower sleeves with a silk cloth.
(2) The friction between the pointer and the dial, the friction between the dial, and the friction between the large pointer and the speed pointer.
(3) The upper and lower sleeves have slightly different axes. Gently tap the measuring cap on the wooden surface. If it is better than before, it indicates that the direction is correct. If it is worse, it indicates that the direction is not correct and another part should be replaced.
(4) There is dirt.
(5) Serious jamming: The deformation of the watch body causes the upper and lower sleeves to have different axes. Remove the movement, apply force to the vise in the circumferential direction on the watch body, and then push the measuring rod. If there is any improvement, it indicates that the direction is correct. Continue to apply force for a period of time, and then remove the watch body; If it's worse, it means the direction is not correct and you need to search for the location again.
3. Deformation of the movement cover plate
The column supporting the cover plate of the movement is relatively high, and the slight deformation of the cover plate causes displacement of the upper and lower bearing holes, thereby affecting the flexibility of the movement transmission. Replace the deformed cover plate of the movement.
4. The balance spring of the ROK dial gauge is not pre tightened or pre tightened
The balance spring of the dial gauge is not pre tightened or pre tightened enough, and after pushing the measuring rod, the movement does not rotate at all or rotates flexibly. The solution is to pre tighten the balance spring. The method of pre tightening the balance spring is to remove the cover plate and circular seat plate of the movement. Remove the small tension spring and force measuring tension spring, and turn the sector gear to the left until it disengages from the shaft gear. Gently press the large pointer shaft and rotate it in the direction of pre tightening the balance spring. The balance spring gear rotates one and a half or two turns, and then turn the sector gear to the right until it meshes with the shaft gear. After meshing, loosen the large pointer shaft, and then
Install other components.
3、 Reasons for the instability of indication values
1. Insufficient pre tension of the balance spring. Loose and disordered inner and outer piles. This is the main reason for the stability of the indication
2. Loose shaft and sleeve, loose needle plate and needle sleeve
3. Loose adjustable column pin
4. Loose end block
4、 Adjustment of indication error
If all items except for the indication error are qualified, the reason for the indication error exceeding the tolerance may be the principle error, wear error, assembly quality, etc. of the lever mechanism.
1. Adjust the length of the lever short arm
The transmission amplification ratio of the micrometer lever mechanism. Inverse ratio. The indication error caused by amplification ratio error. Its error curve is linear or basically linear. It gradually increases or decreases from zero position. When the error value increases, the reason is. The size is short and produces positive errors, which should be increased
Long direction adjustment. When the error value decreases, due to. The negative error caused by the long size should be adjusted in the direction of shortening.
2. Wear and tear of transmission components
Take a small angle of rotation to avoid wear or replace components.
3. Assembly quality
Strive to install each component in the best possible position.