The Testomat ECO hardness analyzer is a simplified version of the 2000 model, which can only measure hardness parameters. It has a single channel measurement, standard 4-20mA output, simplified input-output control functions, and no process display on the panel. Other parameter functions are the same as the Testomat 2000.
Single channel hardness testerTestomat ECO
Measurement principle:
sameTestomat 2000
Measurement range and reagent selection
|
parameter
|
reagent
|
0dH
(Resolution)
|
0f
(Resolution)
|
ppm CaCO3(Resolution)
|
mmol/l
(Resolution)
|
|
water hardness
|
TH2005
|
0.05 – 0.50(0.01)
|
0.09 – 0.89
(0.02)
|
0.89 – 8.93
(0.2)
|
0.01 – 0.09
(0.01)
|
|
water hardness
|
TH2025
|
0.25 – 2.50
(0.05)
|
0,45 – 4.48
(0.1)
|
4,47 – 44.7
(0.9)
|
0.04 – 0.45
(0.01)
|
|
water hardness
|
TH2100
|
1.0 – 10.0
(0.2)
|
1.79 – 17.9
(0.4)
|
17.9 – 179
(3.8)
|
0.18 – 1.79
(0.04)
|
|
water hardness
|
TH2250
|
2.5 – 25.0
(0.2)
|
4.48 – 44.8
(0.4)
|
44.8 – 448
(3.8)
|
0.45 – 4.48
(0.04)
|
System Connection Diagram
Pipeline wiring diagram
sameTestomat 2000
Electrical wiring diagram
|
terminal number
|
terminal
|
function
|
notes
|
|
-
|
PE
|
grounding
|
suitable115/230V !
|
|
--
|
N、L
|
power supply
|
power supply24V/115V/230-240VA,max.4A
|
|
1 2
|
wash
|
Connect the external flushing valve
|
Passive contact output, * high load240VAC, 4 A
|
|
9 10 11
|
LV1
|
Limit contact output1
|
Passive contact output, * high load240VAC, 4 A
|
|
12 13 14
|
LV2
|
Limit contact output2
|
Passive contact output, * high load240VAC, 4 A
|
|
15 16 17
|
Alarm
|
Fault alarm output
|
Passive contact output, * high load240VAC, 4 A
|
|
18 19
|
Stop2
|
External stop analysis contact
|
Passive normally open or normally closed contact input (optional)
|
|
20 21
|
IN2
|
Water meter input contact
|
Only passive normally open contact input
|
|
22 23
|
OUT
|
0/4 - 20mAoutput
|
22=+, 23=-
|
|
24
|
+
|
+12Voutput
|
*High current20mA
|
Technical Parameter
sameTestomat 2000