Welcome Customer !

Membership

Help

Shanghai Xuanren Environmental Protection Instrument Co., Ltd
Custom manufacturer

Main Products:

intelligent-mfg>Products

Shanghai Xuanren Environmental Protection Instrument Co., Ltd

  • E-mail

    sales@sseic.com

  • Phone

    13701858271

  • Address

    Room 103, Building 1, Lingang Excellence Science and Technology Park, No. 877 Jiuxin Road, Songjiang District, Shanghai

Contact Now

Single channel hardness tester

NegotiableUpdate on 06/29
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

The Testomat ECO hardness analyzer is a simplified version of the 2000 model, which can only measure hardness parameters. It has a single channel measurement, standard 4-20mA output, simplified input-output control functions, and no process display on the panel. Other parameter functions are the same as the Testomat 2000.

Product Details

Single channel hardness testerTestomat ECO

Measurement principle:

sameTestomat 2000

Measurement range and reagent selection

parameter

reagent

0dH
(Resolution)

0f
(Resolution)

ppm CaCO3(Resolution)

mmol/l
(Resolution)

water hardness

TH2005

0.05 – 0.50(0.01)

0.09 – 0.89

(0.02)

0.89 – 8.93

(0.2)

0.01 – 0.09

(0.01)

water hardness

TH2025

0.25 – 2.50

(0.05)

0,45 – 4.48

(0.1)

4,47 – 44.7

(0.9)

0.04 – 0.45

(0.01)

water hardness

TH2100

1.0 – 10.0

(0.2)

1.79 – 17.9

(0.4)

17.9 – 179

(3.8)

0.18 – 1.79

(0.04)

water hardness

TH2250

2.5 – 25.0

(0.2)

4.48 – 44.8

(0.4)

44.8 – 448

(3.8)

0.45 – 4.48

(0.04)

System Connection Diagram

Pipeline wiring diagram

sameTestomat 2000

Electrical wiring diagram

terminal number

terminal

function

notes

-

PE

grounding

suitable115/230V !

--

N、L

power supply

power supply24V/115V/230-240VA,max.4A

1 2

wash

Connect the external flushing valve

Passive contact output, * high load240VAC, 4 A

9 10 11

LV1

Limit contact output1

Passive contact output, * high load240VAC, 4 A

12 13 14

LV2

Limit contact output2

Passive contact output, * high load240VAC, 4 A

15 16 17

Alarm

Fault alarm output

Passive contact output, * high load240VAC, 4 A

18 19

Stop2

External stop analysis contact

Passive normally open or normally closed contact input (optional)

20 21

IN2

Water meter input contact

Only passive normally open contact input

22 23

OUT

0/4 - 20mAoutput

22=+, 23=-

24

+

+12Voutput

*High current20mA

Technical Parameter

sameTestomat 2000