Phone
13916388882
-
Address
Room 4005, Block C, Xinhua Garden, 783 Yanchang Middle Road, Shanghai
Shanghai Xufeng Scientific Instrument Co., Ltd
13916388882
Room 4005, Block C, Xinhua Garden, 783 Yanchang Middle Road, Shanghai
22130-2 Liquefied Gas Residue Analyzer
Compliance method:
ASTM D1837; IP 2158; IP 317; SY/T 7509
Scope of application:
This standard is applicable to the determination of residual content of liquefied petroleum gas after volatilization at 38 ℃
Method summary:
Place a 10 mL liquefied petroleum gas sample in a centrifuge tube to evaporate, measure and record the volume of residue left at 38 ℃, and record the phenomenon of adding a certain amount of solvent and residue mixture droplets onto filter paper.
The technical specifications of the instrument are as follows:
Centrifuge tube: 100mL, with a scale of 0.05 mL within the range of 0 mL to 0.5mL
The cooler is made by winding copper pipes with a length of 6m and a diameter of 5mm-7mm into an outer diameter of 63.5mm+1.5nu
The coil is installed in a suitable cold bath
The temperature of the water bath is controlled at 38C+/-2C.