Add buffer solution to the sample, and then add DPD (N, N-diethyl-phenylenediamine) reagent to generate a red complex. Measure the absorbance at 524 nm wavelength and calculate the concentration of free residual chlorine. Under the catalysis of iodide, various forms of chemical residual chlorine (such as monochloramine, dichloroamine, trichloroamine, etc.) can also form red complexes with DPD reagents, and the concentration of free total chlorine is measured at this time. Chlorine dioxide: Chlorine dioxide in water reacts with DPD to produce a pink color, with 20% of the chlorine in chlorine dioxide converted to chlorite. The color reaction is proportional to the chlorine dioxide content in water and is quantified by colorimetry at a wavelength of 524 nm. Glycine converts chloride ions in water into chloroaminoacetic acid without interfering with the determination of chlorine dioxide.
Product features:
★Ultra long measurement interval time, cost-effective maintenance
★Ultra low concentration detection lower limit@50Mm colorimetric dish
★Fully automatic analysis, automatic calibration, and automatic cleaning
★Optional economic typeminiCAP 2000Cl2
★7Inch color touch screen Chinese menu, user-friendly human-machine interface
★Ultra large data storage
★Administrator password protection
★USB software updates and data downloads
Technical indicators:
★ Measurement method: DPD spectrophotometric method
★ Spectrophotometer: Dual path photometer
★ Measurement type: manual or automatic interval analysis
★ Measurement interval: programmable
★ Measurement range:0-1.00 /10.0 mg/l
★ Repeatability: better than 3%
★ Detection limit: 0.005mg/l
★ Measurement time: 5 minutes
★ Display: 7-inch color touch screen
★ Operation menu: Chinese, other options available
★ Data storage: 2G large capacity storage space
★ Digital communication: RS 232/485
★ Analog output: 2-channel 4-20 mA output
★ Protection level: IP 65
★ Power supply: 220VAC, 50/60Hz
★ Size: 600 * 500 * 300 (H * W * D)