E-mail
1450014571@qq.com
-
Phone
18522430527
-
Address
No. 8 Tianmu Industrial Park, Beichen District, Tianjin
Jingqiang (Tianjin) Experimental Instrument Technology Co., Ltd
1450014571@qq.com
18522430527
No. 8 Tianmu Industrial Park, Beichen District, Tianjin
Pressurized Aging VesselThe system design is based on the "Pressure Aging Vessel Accelerated Asphalt Aging Test Method (PAV)" proposed by China Petroleum&Chemical Corporation and drafted by the Heavy Oil Research Institute of the University of Petroleum (East China), as well as the ASTM D6521 and AASHTO PPI test methods. This system is a specialized equipment for evaluating the natural environmental aging status of asphalt.
1. How to translateMain technical features
1. The PAV-1 asphalt pressure aging system consists of a PAV-1-1 asphalt pressure aging chamber and a PAV-1-2 asphalt aging vacuum chamber.
2. The manufacturing, installation, and use of this system comply with the relevant provisions of the Regulations on Safety Supervision of Special Equipment.
3. All technical indicators of this system meet the requirements of similar imported equipment, making it an ideal substitute for imported equipment.
IIMain technical parameters and indicators
(1) PAV-1-1 Asphalt Pressure Aging Chamber
1. Power supply
Power supply voltage: AC220V ± 10% 50Hz
Rated power: 1100W
Rated current of fuse: 10A
2. Pressure vessel
Working volume: 13L
Working medium: Air
Rated working temperature: 110 ℃± 5 ℃
Rated working pressure: 2.1MPa
Design pressure: 2.2 MPa
Hydraulic test pressure: 2.75 MPa
Safety valve opening pressure: 2.15 MPa
Safe bursting disc bursting pressure: 2.6 MPa
3. Pressure controller
Model: WP962-2 (WP-D90)
Accuracy: ± 0.01MPa
4. Temperature and time controller WP962-1 (WP-D90)
Accuracy: ± 0.1 ℃
Time slot: 64 × 1080 minutes
Power off hold function: Yes
5. Dimensions: 630mm × 450mm × 500mm (length, width, height)
(2) PAV-1-2 Asphalt Aging Vacuum Box
1. Power supply
Power supply voltage: AC220V ± 10% 50Hz
Rated power: 900W
Rated current of fuse: 10A
2. Vacuum container
Working volume: 13L
Working medium: Air
Rated working temperature: 170 ℃± 5 ℃
Rated working pressure: Absolute pressure 15kPa
Pressure control method: Electric contact vacuum pressure gauge
3. Temperature and Time Controller XMT-E55T2
Temperature control accuracy: ± 0.1 ℃
Time control method: button
4. Dimensions 560mm × 390mm × 350mm (length, width, height)