Oxidation Stability Tester SCYH8100

Product Details

m1_img(26).png

The SCYH8100 oxidation stability tester is designed and manufactured according to the SH/T0206 and NB/SH/T 0811-2010 "Determination of Oxidation Stability of Unused Hydrocarbon Insulation Oil" standards. This instrument is suitable for determining the oxidation stability of unused, untreated, and inhibitor added mineral insulating oils. Mainly used in the fields of petroleum, chemical, power, pharmaceuticals, commercial inspection, scientific research, environmental protection, etc.

Instrument characteristics
◆ Adopting metal bath heating method, environmentally friendly, pollution-free, simple and easy to use
Adopting PID self-tuning temperature control technology, it heats up quickly and has high temperature control accuracy
The instrument is equipped with an over temperature protection device to prevent loss caused by uncontrolled heating
By setting a preset time parameter, the experiment will automatically stop upon completion, without the need for manual supervision
◆ Real time detection of gas flow using Honeywell quality flow sensors, with high detection accuracy
◆ High definition color screen, full screen touch buttons, Chinese/English interface (optional).
Embedded Linux operating system, SQL ite database technology.
◆ Supports USB interface and audio playback
◆ Supports WIFI connection for remote interactive operation

Technical Parameter
◆ Temperature control range: Room temperature~150 ℃
Temperature control error: ± 0.2 ℃
Flow regulation range: 0ml/min~8ml/min, adjustable arbitrarily
Flow error: ± 0.2ml/min
◆ Sample quantity: 7 channels
Experimental time: 1-500h, continuously adjustable
◆ Minimum adjustable time interval: 1h
Timing error: ≤ 1s/h
Power supply voltage: AC 220V ± 20%, 50HZ ± 10%
Total power consumption: ≤ 1000W
Applicable ambient temperature: 5 ℃ -40 ℃
◆ Applicable environmental humidity: ≤ 85%
◆ Dimensions: 535 * 485 * 523 (instrument), 450 * 120 * 265 (bracket)