The nitrate nitrogen analyzer can be directly connected through a digital communication interface. The NO3-N analyzer provides simple, reliable, and cost-effective on-site sensor monitoring data, and has remote status display, calibration, setting, and fault diagnosis functions. The nitrate nitrogen analyzer can be directly installed in water quality analysis instruments for applications such as reservoirs, rivers, tap water, and sewage, using an IP68 rated dust-proof and waterproof sturdy housing, a 1500N high-strength Kevlar cable, and a transmission distance of up to 1.2 kilometers.
characteristic
- Integrated micro chassis, built-in communicator and sensor
- IP68 sturdy submerged anti-corrosion shell, no control box required, can be directly installed on site
- Plug and play testing, real-time online measurement
- Ultra low power consumption, suitable for outdoor environments without electricity
- 1500N high tensile strength Kevlar cable
- Temperature and chloride ion compensation
- Both the power supply and communication are equipped with surge isolation protection
- RS485 digital communication, can be connected in series to form a loop, saving wiring and engineering costs
- Standard Modbus RTU communication protocol, can be directly connected to PLC and human-machine interface
- Good long-term stability, no need for frequent calibration
- The electrode is protected by a scratch resistant metal film
- The electrode has a maximum lifespan of 12 months or more
- Choose an automatic cleaning brush to reduce maintenance work
- Built in memory that can store calibration parameters, technicians can directly take them outdoors for installation after calibration in the laboratory
- Read monitoring values through AQCFG tool software for calibration, parameter setting, and fault diagnosis
- IECEx CoC/ATEX Cat. 1, Ex ia IIB T5 Ga explosion-proof certification
Typical applications
Drinking water, water treatment, sewage, aquaculture
technical specifications
general
- RS485 digital interface (standard Modbus RTU communication protocol), 19200 bps, 8 data bits, no parity, 1 stop bit; 4~20 mA (optional)
- Data resolution: 16 bits (0.001% FS)
- Surge protection: 4000 V DC
- Power supply: 5~12 V DC; 300 mA
- Protection: polarity, overload, short circuit
- Safety regulations: CE, FCC
measuring principle
Using ion selective electrode method, a thin film on the probe head is used to allow only specific ions to penetrate, generating electrophoretic effect and electrochemical potential. According to the Nernst equation, the potential difference is converted into the concentration of the measured substance. Combine nitrate nitrogen and chloride ion selective electrodes, reference electrodes, and temperature sensors to form a column, and mutually correct and compensate for interference.
Installation method
Immersive, flowing, and plug-in
SMR21Nitrate Nitrogen (NO3-N) Analyzer
| type |
ion selective electrode |
| measuring range |
0.1~1,000 mg/L |
| accuracy |
± 5% reading ± 0.5 mg/L |
| resolution |
0.01 mg/L |
| repeatability |
± 3% reading ± 0.5 mg/L |
| Operating pH |
pH 4~11 |
| Operating Pressure |
Maximum 0.2 Kgf/cm2 |
| operate temperature |
0 ~ 40 °C |
| Water sample flow rate |
0.1~4 m/s |
| reaction time |
2 points (90%) |
| Protection level |
IP68 |
| temperature |
|
| SENSOR |
Pt1000 |
| measuring range |
0~40 °C |
| accuracy |
±0.1 °C |
| resolution |
0.01 °C |
| repeatability |
0.1 °C |
| Chloride ion |
|
| measuring range |
0.1 ~ 1,000 mg/L |
| accuracy |
± 5% reading ± 0.5 mg/L |
| resolution |
0.01 mg/L |
| repeatability |
± 3% reading ± 0.5 mg/L |
| Connection method |
1 "-11PT; 5-core M16 cable head |
| Shell material |
POM |
| size |
Φ 47x279 mm |
| weight |
Nitrate nitrogen analyzer; About 1.5 Kg Nitrate nitrogen and chloride ion analyzer: approximately 1.8 Kg Cable: approximately 80 g/m |