E-mail
jsksdq@126.com
-
Phone
13852565969
-
Address
Chengnan Industrial Concentration Zone, Baoying County, Yangzhou City, Jiangsu Province
Jiangsu Keshuo Electric Co., Ltd
jsksdq@126.com
13852565969
Chengnan Industrial Concentration Zone, Baoying County, Yangzhou City, Jiangsu Province
High voltage dielectric losstest deviceUsed for on-site anti-interference dielectric loss measurement or laboratory precision dielectric loss measurement. The instrument is an integrated structure with built-in dielectric bridge, variable frequency power supply, test transformer, and standard capacitor. Adopting frequency conversion anti-interference and Fourier transform digital filtering technology, the measurement is fully automated and intelligent, and the measurement data is very stable under strong interference. The measurement results are displayed on a large LCD screen and come with a built-in micro printer for printing and output.
1.1 Main technical indicators
1.1.1Measurement of dielectric loss and capacitance
Accuracy: Cx: ±(Reading)×1%+1pF)
tgδ: ±(Reading)×1%+0.00040)
Anti interference indicators: Frequency conversion anti-interference, in200%Can still achieve the above accuracy under interference
Range of capacitance: insideShiHigh voltage: 3pF~60000pF/10kV 60pF~1.2μF/0.5kV
outsideShiHigh voltage: 3pF~1.5μF/10kV 60pF~30μF/0.5kV
resolution: highest0.001pF,4Significant digits
tgδScope: Unlimited, resolution0.001%Automatically identify three types of test samples: capacitors, inductors, and resistors.
Test current range:10μA~5A
Apply high pressure internally: Set voltage range: 0.5~10kV
Maximum output current: 200mA
Boosting and lowering method: Continuous smooth adjustment
Voltage accuracy: ±(1.5%×reading+10V)
Voltage resolution: 1V
Test frequency: 45~65HzInteger frequency
49/51Hz、45/55Hz、47.5/52.5Hzauto frequency conversion
Frequency accuracy:±0.01Hz
External high pressure: Maximum test current during positive wiring1A /40~70Hz
Maximum test current during reverse wiring10kV / 1A /40~70Hz
Measurement time: about30sRelated to the measurement method
1.1.2Other indicators
Input power supply: 180V~270VAC,50Hz/60Hz±1%Powered by mains electricity or generators
Computer interface: standardRS232interface
printer bring your ownMicro thermal printer
Ambient temperature -10℃~50℃
relative humidity: <90%No condensation


Multi functional insulation dielectric loss testeroperating steps:
Ensure that the measuring instrument is placed in a stable and well ventilated environment, away from high temperatures and open flames.
Check if the power connection of the equipment is normal and ensure that the instrument is intact and undamaged.
Check if all connecting wires, plugs, sensors, etc. are intact.
According to the instrument manual, set test parameters such as test voltage and test frequency.
Start the measuring instrument, and the instrument will automatically apply voltage and perform measurements.
Observe the instrument screen and monitor the dielectric loss factor (tan δ) and other relevant parameters.
Record measurement results, including sample information, test voltage, test frequency, tan δ value, etc