MX602 standard gas diluter

NegotiableUpdate on 05/08
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
The MX602 diluter uses a mass flow controller to accurately measure gas flow for linear dilution and calibration.
Product Details

The standard gas diluter provides a standard device for establishing measurement standards for gas analyzers, combustible gases, toxic and harmful gases, and other instruments used in various industries such as metrology, environmental protection, petrochemicals, medicine and health, disease control centers, and coal mines. It prepares standard gases for the calibration, calibration, and testing of these instruments; Used to evaluate the various properties of gas measuring instruments, measuring instruments, and gas sensors, and to prepare standard gases; Reference standards used for various gas analysis methods. The MX602 diluter uses a mass flow controller to accurately measure gas flow for linear dilution and calibration.

Technical parameters

Mass flow controller for dilution gas

05L/min (optional: 10L/min, 20L/min)

Source gas mass flow controller

0-100ml/min (optional: 50ml/min, 1L/min)

accuracy

±1.0% S.P.

dilution ratio

3.5:1950:1 (standard), optionalDilution ratio of 10000 times

Traffic repeatability

<0.5%

Gas distribution uncertainty

<1%

Warm-up time

30min

AUTO ZERO

Automatically adjust the zero point of MFC after preheating is completed

Source gas import

Specification 18 'each

Dilution gas inlet

Specification 14 'one

Standard gas outlet

Specification 18 'each

Bypass outlet

Specification 14 'one

operating display

LCD large screen, built-in software, touch screen operation

Instrument power supply

220VAC

Usage environment

5℃45℃

External dimensions

427-380-185mm

Instrument weight

About 5kg

Optional accessories:

Inert options for the entire flow system: Pipelines, joints, and components through which standard gas passes are all treated with silane, which can meet the dilution of polar VOCs such as sulfides.