High temperature test chamber room temperature+45 ℃~350 ℃

NegotiableUpdate on 01/09
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
High temperature test chamber (floor type - LTV) with room temperature+45 ℃~350 ℃/room temperature+113 ℉~662 ℉ -70~180/-94~356. The temperature of all models has been tested and meets DIN 12880 and IEC 60068-3-5 standards. Model: LTV-012/025/040/070/100
Product Details
model LTV-012 LTV-025 LTV-040 LTV-070 LTV-100

Chamber volume(L / cu ft)

125 / 4.4

253 / 8.9

420 / 14.8

720 / 25.4

1000 / 35.3

controller

Programmable, touch screen controller(3.5 "color display screen)

temperature range( ℃ / ℉)

When the ventilation opening is closed

Room temperature+45~350/Room temperature+81~662

When the ventilation opening is closed

Up to 115/up to 239

Up to 115/up to 239

Max 180/Max 356

Up to 150/up to 302

Maximum 210/Maximum 410

Temperature variation1) 2)
At 100 ℃( ±℃ / ℉)

0.4 / 0.72

0.4 / 0.72

0.3 / 0.54

0.5 / 0.90

0.4 / 0.72

Temperature fluctuations2)

( ±℃ / ℉)

At 100 ℃

0.5 / 0.90

0.5 / 0.90

0.6 / 1.08

0.7 / 1.26

0.4 / 0.72

At 150 ℃

0.7 / 1.26

0.7 / 1.26

1.0 / 1.80

1.1 / 1.98

0.6 / 1.08

At 200 ℃

1.1 / 1.98

1.1 / 1.98

1.7 / 3.06

1.9 / 3.42

1.1 / 1.98

At 250 ℃

1.6 / 2.88

1.6 / 2.88

2.5 / 4.5

2.7 / 4.86

1.6 / 2.88

At 300 ℃

2.4 / 4.32

2.4 / 4.32

3.4 / 6.12

3.4 / 6.12

2.4 / 4.32

At 350 ℃

3.4 / 6.12

3.4 / 6.12

3.4 / 6.12

4.1 / 7.38

3.2 / 5.76

Heating time2) (minutes)
From room temperature to 350 ℃

41

41

43

40

45

Temperature recovery time2)
After the door is opened(minutes)

At 100 ℃

8.5

8.5

6

6

7.5

At 150 ℃

8.5

8.5

9

7.5

9.5

At 200 ℃

3.5

3.5

8.5

8.5

8.5

At 250 ℃

5

5

11.5

8.5

9.5

At 300 ℃

7.5

7.5

12.5

8.5

9

At 350 ℃

8.5

8.5

12.5

10

10

Ventilation rate(Approximately, x/h)

213

266

173

233

193

Air circulation(Approximately, x/h)

374

590

294

348

220

Gas discharge flow rate(Approximately, L/min)

380

1240

1400

1910

1560

Air flow rate(m/s)

0.2 ~ 0.6

0.2 ~ 0.6

0.2 ~ 0.6

0.2 ~ 0.6

0.2 ~ 0.6

Number of steel wire frames(Standard/Maximum)

2 / 12

2 / 17

2 / 22

2 / 29

2 / 32

Rack spacing(mm / inch)

30 / 1.2

Maximum load-bearing capacity of each shelf(kg / lbs)

15 / 33.1

20 / 44.1

30 / 66.1

40 / 88.2

45 / 99.2

All maximum load-bearing capacity(kg / lbs)

50 / 110.2

70 / 154.3

90 / 198.4

120 / 264.6

150 / 330.7

size
(W×D×H)

internal dimensions

(mm / inch)

500×500×500/

19.7×19.7×19.7

600×650×650/

23.6×25.6×25.6

750×700×800/

29.5×27.6×31.5

900×800×1000/

35.4×31.5×39.4

1000×910×1100/

39.4×35.8×43.3

external dimensions

(mm / inch)

1100×745×1362/

43.3×29.3×53.6

1200×995×1559/

47.2×39.2×61.4

1350×1045×1712/

53.1×41.1×67.4

1500×1145×1912/

59×45.1×75.3

1600×1255×2012/

62.9×49.4×79.2

Net weight

(kg / lbs)

170 / 374.8

260 / 573.2

320 / 705.5

390 / 859.8

440 / 970

communication interface

Standard RS-232 interface(Optional RS-485 interface)

Power requirements(230V, 60Hz) 1ph / 27.6A
Item Number AAHK2011K - - - -
Power requirements(230V, 50Hz) 1ph / 27.6A
Item Number AAHK2012K - - - -
Power requirements(380V, 50Hz) 3ph / 13.9A 3ph / 18.5A 3ph / 23.4A 3ph / 27.9A
Item Number - AAHK2028K AAHK2038K AAHK2048K AAHK2058K

1) Meets the requirements of IEC 60068-3-5.
2) Compliant with DIN 12880 requirements.

The above parameter values were measured at 230V/60Hz and 380V/50Hz
※ Due to constantly updated parameters, please understand if not notified.
※ It is a registered company under the FDA and its products are registered with the FDA.

Performance Characteristics

  • It can have special purposes, such as heat treatment and drying treatment.
  • Microfabrication PID controller/automatic calibration function/automatic adjustment function.
  • The door is securely sealed.
    -Heat resistant silicone can provide a strong sealing effect, preventing heat loss inside the cavity (applicable to LBV series)(Applicable to LBV series)
    -Non conductive and heat-resistant ceramics can provide a strong sealing effect, preventing heat loss inside the cavity(Applicable to LTV series)
    -A 2-node latch can also improve sealing performance.
  • A good insulation system can have better energy utilization efficiency.
    -Even if the system operates at a high maximum temperature, the surface temperature of the equipment will not exceed 51 ℃(Compliant with DIN EN 563 standard)
  • The door has a reliable design.
    -When the door is opened 90 degrees, it can carry a weight of 50kg at a distance of 30cm from the door latch.

Safety features

  • After power outage and restoration, the operator can choose to operate automatically or manually.
  • Over temperature protection.
  • Current overload protection and leakage protectionGround leakage circuit breaker
  • Independent safety fuse.
    -Important parts such as controllers and heaters have short-circuit protection. Especially each heater has an independent fuse, which can prevent all short circuits.
  • Reverse phase protection.
    -The electrical connection error indicator can help users quickly repair connection problems.
  • Motor current overload protection.
    -When the motor current is overloaded, the system will automatically cut off the power supply.
  • Door opening alarm function.
    -When the door is open, the system will automatically turn off the heater and fan.
  • Wire damage alarm function.
    -When the surface of the wire is damaged, it will automatically shut down.
  • Built in door lock and standard key.

Easy to operate

  • The inner cavity is made of corrosion-resistant stainless steel.
  • Efficient utilization of space and convenient placement of samples through internal cavity design.
    -Comes with two steel wire frames as standard, capable of carrying heavy samples.
    -Optional perforated shelf for carrying overweight (40kg)Optional
  • A simple and easy-to-use LCD controller.
    -The control signal of the device can be promptly fed back through the color LCD touch screen. -The control signal of the device can be promptly fed back through the color LCD touch screen.
    -The set temperature value and the current temperature value can be displayed on the screen simultaneously.
    -Support programming functionality.
    -Displaying the operation process in graphical mode makes the data easier to understand.
    -The automatic mode can be turned on or off, and can be programmed to run automatically.
    -The monitor can display abnormal temperature changes or power failures.
    -The interface supports three languages(English/Chinese/Korean)
    -6-channel digital recorder can record temperature and other parametersOptional
  • Computer communication interface.
    -RS-485 communication interfaceOptionalUp to 32 units can be connected. Multiple units can be monitored and operated simultaneously.
    -With this software, programming will become even simpler.
    -Can check the running status in real-time. The data is displayed in graphics, making it easier to understand.
    -It has storage and programming functions.
    -Data can be saved in EXCEL format or document format.
    -It can store program information, set value backups, reset, and temperature values.
    -Standard RS-232 communication interface.