Phone
13911644071
-
Address
6G, Yaoyan Building, Jiulong Mountain Villa, Xihongmen, Daxing District, Beijing
AGC Instruments Ireland Limited Beijing Representative Office
13911644071
6G, Yaoyan Building, Jiulong Mountain Villa, Xihongmen, Daxing District, Beijing
High purity argon chromatographThe ADD argon discharge ionization detector is mainly used for the analysis of H2, O2, N2, CH4, CO, and CO2 in high-purity argon and ultra pure argon. All of its connection interfaces use 1/8'VCR to ensure the airtightness of the instrument. The installed valve has undergone strict leakage testing and is installed in a casing with Ar gas purging, isolated from the atmosphere. Therefore, environmental pollution is eliminated, ensuring that the instrument has a more stable, sensitive, and longer service life.
1、 Compliant with standards:
GB/T 16945-2009 "Gas Argon for Electronic Industry" GB/T 4842-2006 "Argon"
2、 Technical specifications:
1. Detection limit
AGC NovaCHROM2000High purity argon chromatographDetection limit (ppb) | ||||||
component |
H2 |
O2 |
N2 |
CH4 |
CO |
CO2 |
detection limit |
10ppb |
25ppb |
50ppb |
20ppb |
100ppb |
30ppb |
2. Performance indicators:
detector |
ADD high-frequency argon discharge detector |
repetitiveness |
<0.5% of range |
precision |
<0.5% of scale |
response time |
<0.5 seconds (90%) |
Column box temperature |
Environmental temperature 5 ℃~450 ℃ |
Multi column box |
have |
temperature-programmed |
Level 10 |
heating rate |
0.1 ℃~40 ℃/min (increments of 0.1 ℃) |
Cooldown |
300℃~50℃(5min) |
EPC electronic flow control |
have |
power supply |
115/220V,50/60Hz,2300W |
weight |
30kg (depending on configuration) |
size |
572mm(W)×508mm(H)×559mm(D) |
3. Carrier gas purifier:
Purifier type |
Renewable Purifier |
Purification level |
He can be purified from 5.5 grade to 6.5 grade or even higher. |
Cleared impurities: |
H2O, O2, CO, CO2, THC, H2, N2 (each less than 1 ppb) |
Suggested temperature for use: |
350°C |
Maximum operating temperature: |
450°C |
Maximum operating pressure: |
10 bar |
Zui high flow rate: |
500 ml |
Standard installation interface: |
1/8' VCR |
Power requirements: |
110/220/240 V 50/60 Hz , 500 VA |
Net weight: |
9 kg |
4. Workstation performance:
Model: |
AGC Trend-Vision |
manufactor: |
AGC Instruments Ireland Limited |
output: |
Zui has 16 independent 4-20mA outputs Top 21 passive relays with high/low alarms for electric shock |
Communication interface: |
350°C |
resolution |
24 bit resolution |
input range |
1V |
Heavy linearity |
±0.002% of full scale |
Integral sensitivity |
1μv•sec |
System compatibility |
• Windows XP ® • Windows Vista ® • Windows 7 ® • Windows 8 ® • Windows 10 ® |
3、 Technical advantages:
● Core component - AGC's DID HFADD FID TCD detector * Based on independent research and development by AGC in Ireland, it has strong performance, high sensitivity and stability, and is the choice for analyzing low content impurities in high-purity gas;
Dual detector parallel connection - AGC company has developed a dual detector parallel connection system, equipped with two independent detectors. Their temperature control system and operation are both independent detectors, which can be connected in series or parallel;
● Gas path system - unique center cutting and blowback system technology, which can complete the analysis and detection of multiple complex components at once;
● Software System - Trend Vision, specially developed for AGC chromatography. In addition to completing test reports, automatically analyzing/storing data, automatically drawing spectra, controlling gas chromatography, adjusting retention time, and various cutting actions, it can also control up to 4 analytical instruments simultaneously;
● Carrier Gas Purification System - AGC 75-802 Carrier Gas Purifier is a renewable purifier that can purify 5.5N grade gas to 6.5N or higher. The ultra-high purity of the carrier gas gives AGC high-purity gas chromatograph higher sensitivity;
● Carrier gas switching device - The configuration of CGDP-01 carrier gas switching device can simultaneously carry two bottles of carrier gas, avoiding the entry of polluted gas when switching gas sources after the carrier gas is used. The configuration of the carrier gas switching device ensures higher work efficiency and analysis quality of AGC gas chromatography.