- Phone
-
Address
No. 10 Jinye Street, Xihongmen Town, Daxing District, Beijing
Beijing Huakeyi Technology Co., Ltd
No. 10 Jinye Street, Xihongmen Town, Daxing District, Beijing
Total phosphorus refers to the total amount of orthophosphate, pyrophosphate, pyrophosphate, polyphosphate, and organic phosphate in water. The results of measuring the conversion of various forms of phosphorus into orthophosphate after high-temperature digestion of water samples.
The instrument adopts automatic control of water sample collection, high-temperature digestion, reaction measurement, data processing and transmission, and quickly and continuously completes the entire process of total phosphorus analysis in wastewater.
According to the ammonium molybdate spectrophotometric method, in an acidic solution, potassium persulfate is used as a decomposing agent to convert polyphosphate and organic phosphorus into orthophosphate. orthophosphate reacts with ammonium molybdate to form a yellow phosphomolybdate antimony complex, which is then reduced to phosphomolybdate blue by ascorbic acid. The phosphate content is then measured using a total phosphorus analyzer.◆The 7-inch LCD touch screen is easy to operate, with a beautiful and elegant interface, and rich and intuitive display content.
◆The integrated design of digestion and measurement ensures more accurate and rapid measurement results.
◆The light source adopts imported monochromatic photodiodes, which have excellent performance, energy concentration, and low power consumption.
◆Adopting high-precision injection pump, with high measurement accuracy and low reagent consumption.
◆Adopting acid alkali and high temperature resistant valve circuits, with long service life and minimal manual maintenance.
◆The instrument is equipped with a preprocessing device inside, which can meet the requirements of different water quality measurements.
◆It has an automatic calibration function to ensure that the instrument can measure more accurately.
◆Equipped with intelligent maintenance function, it automatically cleans after each inspection to ensure more accurate measurements and reduce manual maintenance.
|
Measurement method: |
Ammonium molybdate spectrophotometric method |
|
Measurement range: |
0.05-50mg/L, expandable range |
|
resolving power: |
0.01mg/L |
|
Indication error: |
≤1mg/L:±0.05mg/L;>1mg/L:±3% |
|
Repeatability: |
<3% |
|
Stability: |
24hWithin ±5% |
|
Measurement interval: |
1-24Hour optional |
|
Measurement method: |
Loop measurement, fixed-point measurement, trigger measurement |
|
Environmental temperature: |
5-45℃ |
|
ambient humidity: |
<95% |
|
Display: |
7Inch LCD touch screen |
|
Language: |
Chinese/English |
|
Output method: |
4-20mA |
|
Communication method: |
RS485,MODBUS |
|
Power Supply: |
AC(220±22)V;50HZ |
|
Power: |
<100W |
|
Appearance dimensions: |
500mm*395mm*1450mm(Long)*wide*High) |
|
Total weight of the instrument: |
make an appointment80Kg |
|
Other functions: |
Automatic calibration, intelligent maintenance, data storage, trend curve, etc |