Instrument Introduction:
The latest Chembet PULSAR TPR/TPD (p/n 02139-1) launched by Kantar Instruments represents the best level of fully automated flow analysis and characterization of catalysts. The new Pulsar building is built on top of the world-renowned ChemBET and combines the fully automatic functions of Autosorb-1C/TCD
Technical parameters:
Performance indicators:
Number of sample stations: 2 (one analysis station and one preparation station)
• Titration quantitative ring: Automatic - provides 50, 100, and 250 µ l quantitative rings (other volumes are optional)
Gas input port: 5 (four automatic switch gas lines for quantitative/titration gas lines,
6 (+carrier gas path (if adding mixer option))
• Sample tube: High temperature chemisorbed U-shaped quartz sample tube, traditional U-shaped quartz tube or boron silicide sample tube (Pyrex) ®)
Sample tube bypass valve: Yes
• Steam trap: Yes, (located between the sample tube and detector to remove impurity gases)
• GC/MS connection methods: 3 types (a: at the back end of the sample tube; b: After the cold trap and before the detector; c: After the detector)
• Volume: from 1x10-3 cm3 to>1x102 cm3
• Minimum specific volume: 1x10-4 cm3/g
Total surface area: 0.1 to 280 m2
Specific surface area: Starting from 0.01 m2/g, the upper limit is determined by the minimum weighing accuracy
• Pore volume ranging from 1x10-4 cm3 to 1.5x10-1 cm3
• Volume accuracy: ± 1%
• Reproducibility: 0.5%
Chemical adsorption gases: typical gases H2, O2, CO, NO, SO2, NH3
• Physically adsorbed gases: typical gases N2, Ar, Kr, CO2 (helium as carrier gas)
Sample tube volume: 1.5cm3 standard tube, other sizes available
Physical adsorption cold trap: generally using liquid nitrogen
• Cold trap refrigerants: LN2/alcohol, dry ice/acetone, liquid nitrogen, etc
• Wetted materials: stainless steel, PTFE, EP, butadiene rubber, glass, rhenium/tungsten, gold
Preparation temperature: Heat the package with quartz to 450 º C
Use a high-temperature furnace to reach up to 1100 º C
TPR/TPD heating rate: 1 º C min-1 to 100oC min-1 (up to 500 º C)
1ºC min-1 to 50ºC min-1 (up to 750ºC)
1ºC min-1 to 30ºC min-1 (up to 1000ºC)
1ºC min-1 to 20ºC min-1 (up to 1100ºC)
Temperature accuracy: ± 1 deg C
Temperature repeatability: ± 1 deg C
High temperature furnace control: through software
Sample thermocouple: inside the sample tube
Forced air cooling system: Yes
Cooling time (from maximum temperature to 100 deg C):<30 minutes
Volume and electrical parameters:
Host volume: 45cm x 63cm x 42cm (WxHxD)
Host weight: 24 kg
Operating temperature: 15 º C to 35 º C
Storage temperature: 0 º C to 50 º C
Humidity range: 20-80% (non condensing)
Voltage: 100/120&220/240 V (ac)
Voltage frequency: 50/60 Hz
Main features:
The latest Chembet PULSAR TPR/TPD (p/n 02139-1) launched by Kantar Instruments represents the best level of fully automated flow analysis and characterization of catalysts. The new Pulsar building is built on top of the world-renowned ChemBET and combines the fully automatic functions of Autosorb-1C/TCD:
- The latest TPRWin PC operation and application software fully automated programmed control analysis sequence
Fully automatic quantitative ring injector and automatic intake valve for pulse titration analysis of metal area and dispersion determination
The temperature climb of the high-temperature furnace can be used for both programmed heating experiments and sample preparation. Both methods can rapidly cool the high-temperature furnace by forcing high-speed air flow.
- Retaining ChemBET's proven TCD detector, it has strong antioxidant and ammonia corrosion resistance capabilities, and stable current control ensures baseline stability and reproducibility of signals
Stainless steel pipelines with maximum chemical compatibility are suitable for a wide range of gases.
The standard high-temperature quartz sample cell can be equipped with a thermocouple to provide accurate temperature measurement of the sample.
- Built in data acquisition card and temperature controller