Product details Product name: HM-20, HM-21 flame spectrophotometer Product description: HM-20, HM-21 are analytical instruments designed and manufactured according to the basic principles of emission spectroscopy, widely used in the analysis and determination of agricultural fertilizers, soil, cement, ceramics and other industries, as well as the silica industry; Pathological research in healthcare
Product Details

product details

Product name: HM-20, HM-21 flame photometer

Product description of flame spectrophotometer:

HM-20 and HM-21 are analytical instruments designed and manufactured based on the basic principles of emission spectroscopy, widely used in industries such as agricultural fertilizers, soil, cement, ceramics, and the silica industry for analysis and determination; Pathological research in healthcare. The flame photometer uses liquefied gas as fuel gas.

Characteristics of flame spectrophotometer instrument:

1) K and Na are displayed simultaneously;

2) Using touch sensitive buttons for easy operation;

3) Equipped with flame size pre selection function;

4) Equipped with flameout protection device;

5) Equipped with RS232 interface, it can connect to printers and computer PC software;

6) Equipped with a moisture separator.

Technical parameters of flame spectrophotometer:

model

HM-20

HM-20 (soil fertilizer specific)

HM-20 (cement specific)

HM-21

Display Range

potassium

0-100

0-50

0-80

0-100

ug/ml

sodium

0-100

0-160

0-160

0-100

Detection limit

potassium

≤0.004

≤0.004

≤0.004

≤0.004

mmol/L

sodium

≤0.008

≤0.008

≤0.008

≤0.008

Linear error

potassium

≤0.005

≤0.005

≤0.005

≤0.005

mmol/L

sodium

≤0.03

≤0.03

≤0.03

≤0.03

stability

Continuous injection of standard solution, with a relative maximum change in instrument readings of ≤ 3% within 15 seconds,

Measure once per minute, for a total of 6 measurements, the relative maximum change in instrument readings is ≤ 15%

response time

<8 seconds

Minimum suction and spray volume

<6mL∕min

Repeatability

Repeat the same standard solution for 7 consecutive independent measurements ≤ 3%

Test elements

Potassium and Sodium

Potassium and Sodium

Potassium and Sodium

Potassium and Sodium

Number of channels

2

2

2

2

Curve calibration

Two points

Line fitting

printer

Optional selection

Connect computer

yes