Brief introduction
EPLEXOR ® The series is a DMA/DMTA testing instrument with a maximum force value of ± 500 N. It can characterize materials under dynamic or static mechanical forces. Its application range covers various materials, including elastomers and polymers, composite materials, metals, glass, ceramics, biomaterials and food, adhesives, and liquids. The main difference between this series of testing instruments lies in their maximum dynamic force range (± 25N, ± 50N, ± 150N, ± 500N).
DMA/DMTA has a modular design that allows measurement in various modes such as tension, compression, bending, and shear.
The instrument can be equipped with different expansion accessories, making it possible to fully expand its application areas in the future.
All testing instruments in this series comply with DIN 53513, ISO 6721/1, ISO 6721/4, ISO 6721/5, ISO 6721/6, ISO 4664, ASTM D4065, Related standards such as ASTM D4473.
Flexible and future oriented
The instrument can be equipped with different types of force and displacement sensors, as well as different types of furnace bodies, which makes it easy to upgrade from the basic system after installation.
Strength value
The instrument supports a maximum static load of 1500 N and a maximum dynamic load of ± 500 N; This is particularly meaningful for the research of thermosetting resins, elastomers, composite materials, metals, glass, or ceramics.
Two independent motor drives
... including a servo motor for static loads and a vibrator for dynamic loads
Replaceable force sensor
... can be easily replaced by the user; The nominal load range is from 10 N to 5000 N
Continuous operation through an automatic sampler
For stretching, compressing, and bending samples, an automatic sampler can work day and night throughout the entire temperature range, making full use of the measurement time at night and on weekends.
Optimize temperature scanning for large samples
Even for large samples with low thermal conductivity (such as large rubber specimens), the instrument can ensure uniform heating.
Highly economical LN2 cooling
Low consumption rate of liquid nitrogen.
Synchronous measurement of dynamic mechanical and dielectric properties of materials
III EPLEXOR ® Series - Software Features
Powerful EPLEXOR ® Software based on the Windows operating system, including data and curve analysis, hysteresis plot analysis, main curve calculation, and more. Simultaneously containing different templates for tensile, compressive, or bending testing.
Software features:
●Frequency scanning, range 0.001 Hz... 100 Hz (optional 0.0001 Hz... 200 Hz)
●Temperature scanning (controllable temperature changes at a fixed frequency)
●Time scanning from 1 second to 10 seconds7s
●Frequency scanning at a certain temperature, with temperature adjusted in a step-by-step manner
●Strain amplitude scanning under dynamic and static forces, with force values adjusted in steps of equidistant or logarithmic division
●Main curves (TTS, WLF, numerical mastering), segmented testing
●Analyze complex modulus (E *, G *), storage modulus (E ', G'), loss modulus (E ", G"), damping factor (tg δ), temperature scan, stress-strain scan, time scan, glass transition temperature; Optional features include creep, relaxation, fatigue, energy loss, hysteresis, Payne/Mullins effect analysis, and crack growth testing
4 EPLEXOR ® Series - Related Attachments
Sample holder suitable for numerous applications
EPLEXOR can be used for various types of materials, from liquids and reinforced resins to metals and ceramics ® Come and conduct research.
Synchronize DMA/DMTA and DEA
Study transport and relaxation processes at the molecular level.
UV Accessories
Study the reaction under UV light irradiation.
Humidity Generator (HYGROMATOR) ®)
An attachment that can be used to study the water absorption process of samples such as plastics and biopolymers. By using a humidity generator, a relative humidity atmosphere ranging from 5% to 95% can be created within a temperature range of 5 ° C to 95 ° C.
Immersion container
... can be measured in tension, compression, and bending modes; Used to study the aging or plasticizer effects caused by the contact of samples with water or oil.
Goodrich Flex Test Module
... used to optimize the thermal properties and lifespan of tire mixtures; Apply cyclic load and test for temperature changes caused by heat accumulation.
Cooling options
When using a standard furnace body (-150... 500 ° C), two different cooling systems can be selected, including liquid nitrogen cooling system (as low as -150 ° C) and air cooling system (as low as -60 ° C).
DMTA measurement in reactive atmosphere
For on-site research, equipment with exhaust can be equipped to protect the laboratory environment.