The adjustable precision PID controller of the cooling water circulation device (high-end low-temperature type) is suitable for various condensation tasks such as scientific, research, industrial laboratories, etc. Model: HS-15/20/25/35/45/45H/55/55H
Product Details

Digital display controller

RS-232 interface, etc

Performance Characteristics

  • Temperature range: -20 ℃~40 ℃
  • The maximum cooling power is 7.1 kW at 20 ℃.
  • The maximum flow rate of the circulating pump is 70L/min. and the pressure is 6.0 bar.
  • Accurate temperature stability without heating: ± 0.2 ℃ (at 15 ℃)
    -Innovative and unique controllers can adjust cooling capacity from 1% to 100%.
  • The cooling system can reduce energy waste and achieve maximum cooling effect.

Easy to operate

  • Simplified VFD touch screen for operation.
    -Set the resolution to 0.1 ℃/℉
    -Quick keyboard lock function to prevent unauthorized operations.
  • The LED liquid level indicator can also be seen from a distance.
  • Optional external temperature sensor for more precise temperature control.
  • Built in pressure gauge, which can clearly show the pressure of the pump at a glance.
  • The high/low pressure gauge can clearly check the status of the refrigeration system at a glance.(Applicable to H series)
  • The spacious inlet allows for easy and safe pouring of the solution into the system.
  • Two backup valves can prevent coolant leakage.
  • Mobile grid vents can reduce contamination of the condenser.
  • Standard casters for easy installation and movement.
  • The RS-232 interface can be used for external control and data collection.

Safety features

  • Automatic diagnostic function, automatically identifying errors.
  • Safety protection function with signal alarm.
    -Low liquid level and no bath protection, there will be an alarm sound to remind when there is an error signal on the display.
    -Over temperature protection, if the device experiences over temperature due to external environment, an audible and visual alarm reminder signal will appear.
    -Current overload protection.
  • The environmentally friendly R-404A/R-507 coolant can prevent environmental pollution.
  • The keyboard can prevent splashing with shower gel
    model HS-15 HS-20 HS-25 HS-35 HS-45 HS-55 HS-45H2) HS-55H2)

    Temperature1)

    (Bath solution: ethylene glycol and water ratio: 1:1)

    Temperature range( ℃ / ℉) -20 ~ 40 / -4 ~ 104

    Temperature stability, at 15 ℃,

    ( ±℃ / ℉)

    0.2 / 0.36

    refrigeration power

    (Bath solution: ethylene glycol and water ratio: 1:1)

    At 20 ℃(kw) 1.45 1.8 2.5 3.3 6.5 7.1 6.5 7.1
    At 10 ℃(kw) 1.15 1.5 1.8 2.3 4.5 6 4.5 6
    At 0 ℃(kw) 0.86 1.15 1.1 1.7 3 4.1 3 4.1
    At -10 ℃(kw) 0.62 0.85 0.65 1 2.1 2.5 2.1 2.5
    At -20 ℃(kw) 0.3 0.4 0.4 0.55 1.2 1.5 1.2 1.5
    pump Maximum flow rate(L / min, gal / min) 40 / 10.57 70 / 18.49
    Maximum pressure(bar / psi) 3.3 / 47.86 6 / 87
    size Maximum filling volume(L, cu ft) 13.5 / 0.47 25 / 0.88 39 /1.37
    Pipe diameter(mm / inch) 20 / 3/4
    Infusion import(Ø, mm / inch) 120 / 4.7

    external dimensions

    (W×L×H, mm / inch)

    515×715× 835

    / 20.3×28.1×32.9

    550×900×1140

    / 21.7×35.4×44.9

    605×1045×1300

    / 23.8×41.4×51.2

    Net weight(kg / Ibs)

    91.8

    / 202.4

    92.65

    / 204.3

    141.3

    / 310.9

    146.3

    / 321.9

    168 ±10

    / 370 ±22

    173 ±10

    / 381 ±22

    171 ±10

    / 377 ±22

    176 ±10

    / 388 ±22

    Power requirements(230V, 60Hz) 6.5 A 7.5 A 9.5 A 13.5 A
    Item Number

    AAH

    66011K

    AAH

    66021K

    AAH

    66031K

    AAH6

    6041K

    - - - -
    Power requirements(230V, 50Hz) 6 A 7 A 7.5 A 12.5 A
    Item Number

    AAH

    66012K

    AAH6

    6022K

    AAH

    66032K

    AAH

    66042K

    - - - -
    Power requirements(380V, 60Hz) 5.5 A 7 A 5.5 A 7 A
    Item Number - - - -

    AAH

    66259K

    AAH

    66469K

    AAH

    66359K

    AAH

    66569K

    Power requirements(380V, 50Hz) 5 A 6 A 5 A 6 A
    Item Number - - - -

    AAH

    66258K

    AAH

    66468K

    AAH

    66358K

    AAH

    66568K

    1) The technical parameters comply with DIN 12876 standard.