BINDER VDL23 vacuum drying oven
1、 Product Introduction:
1. Temperature range: ambient temperature plus 15 ℃ to 200 ℃
2. Equipment chambers that comply with ATEX standards: Ex II -/3G c IIB T3-T1 Gc X
3 APT.lineTMPreheating chamber technology
4. Controller using interval programming and real-time programming
5. 2 aluminum expansion brackets
6. Inert gas interface
7. Anti impact safety glass panel with spring plates
8. Overpressure sealed toolbox with compressed air interface
9. Pressure monitor, used for heating and releasing<125 mbar
10. Fireproof sealing ring
11. Independent adjustable temperature safety device level 2 (DIN 12880) with optical alarm
12. Computer interface: RS 422
2、 Technical parameters:
| Product model | VDL023-230V1 |
| Choose a model | standard |
| Item code | 9030-0038 |
| measure | |
| Internal volume [L] | 23 |
| Net weight of equipment (unloaded) [kg] | 63 |
| Maximum total load [kg] | 35 |
| Maximum load per partition [kg] | 20 |
| The dimensions of the casing do not include additional components and connectors | |
| Net weight width [mm] | 515 |
| Net weight at height [mm] | 655 |
| Depth net weight [mm] | 500 |
| Wall to side distance [mm] | 135 |
| Wall distance from rear [mm] | 100 |
| Observation window width [mm] | 250 |
| Observation window height [mm] | 207 |
| Internal dimensions | |
| Width [mm] | 285 |
| Height [mm] | 285 |
| Depth [mm] | 295 |
| The number of doors | |
| Outer door | 1 |
| firmware | |
| Number of partitions (hours/maximum) | 2/4 |
| Temperature performance data | |
| The temperature range exceeds the indoor temperature by 15 ℃ to [℃] | 200 |
| Temperature deviation within 100 ℃ [± K] | 1.5 |
| Temperature fluctuation [± K] | 0.1 |
| Heating time to 100 ℃ [min] | 65 |
| Vacuum data | |
| Allowable maximum vacuum [mbar] | 0.01 |
| Leakage rate [bar/h] | 0.01 |
| Electrical data | |
| Rated voltage [V] | 230 |
| Power frequency [Hz] | 50/60 |
| Rated power [kW] | 0.8 |
| Equipment safety device [A] | 10 |
| Phase (rated voltage) | 1~ |
| Connecting components | |
| Vacuum interface with small flange | 16 |
| Test hole with small flange | 16 |
| Inert gas interface with flow restrictor (RP) | 3/8 |
| Compressed air interface for overpressure sealing structure [mm] | 8 |
1. All technical data is only applicable to standard specifications of unloaded equipment at an ambient temperature of 22 ± 3 ℃ and a power supply voltage fluctuation of ± 10%. Determine temperature data according to BINDER factory standards and DIN 12880:2007, and adapt to the recommended wall gap of 10% of the internal chamber height, width, and depth. For series devices, all data are typical averages. We reserve the right to make technical changes.