The AT-10TVP Injector comes with the following standard accessories:
Power connection cable (EU standard)
Serial interface (RS-232)
1/16 tube, inner diameter 0.50mm, 60cm (SS or PEEK)
1/16 pipe sleeve, stainless steel
Waste liquid bottle (1000ml)
Cleaning bottle (500ml)
Operation Manual
Fuse, 2A
1. Disk sampling
The AT-10TVP PROJECT automatic sampling system has the technical design feature of mechanical disc sampling, which can achieve a longer service life. Automatic lubrication bearings can minimize daily maintenance and avoid trouble caused by dust environments.
The high-precision stepper motor shaft can be accurately positioned, and the injection dose is provided using a micro mode, which can accurately determine the position of the sample bottle with high resolution.
2. Simple operation, easy to understand
The injection valve and injection port with a quantitative ring can be directly accessed from the chamber door on the instrument side panel without removing any protective cover. The replacement of the suction needle does not require any necessary tools.
Technical indicators:
(1) Repeatability: RSD<0.5% (full quantification tube injection)
(2) Linearity:>0.999 (full quantitative tube injection)
(3) Residue: < 0.02%
(4) Number of sample bottles: 49 2mL sample bottles, 1 20mL cleaning waste bottle;
(5) Syringe (pump) specification: Injection volume 250uL
(6) Electric injection valve speed: < 100mS
(7) Position control accuracy: < 0.5mm
(8) Injection motion mode: The sample disk moves in a circular motion, and the injection needle and injection pump move up and down;
(9) Injection mode: no cleaning injection, cleaning after injection, cleaning before injection, cleaning before and after injection;
(10) Injection needle cleaning method: inner wall
(11) Communication method: RS232
(12) Power supply: AC220V 50Hz; Power 150W