E-mail
1375312008@qq.com
-
Phone
18121126516(微信同号)13301881771
-
Address
Building 10, 4th Floor, Block F, No. 1188 Lianhang Road, Minhang District, Shanghai
Shanghai Yudong Electronic Technology Co., Ltd
1375312008@qq.com
18121126516(微信同号)13301881771
Building 10, 4th Floor, Block F, No. 1188 Lianhang Road, Minhang District, Shanghai

| Performance |
MeasuringRange |
0.5 to 1 ppm |
1 to 20 ppm |
2 to 30 ppm |
30 to 60 ppm |
Number of PumpStrokes |
8 |
4 |
2 |
1 |
CorrectionFactor |
1/4 |
1/2 |
1 |
2 |
SamplingTime |
1 minute per pumpstroke |
|||
DetectingLimit |
0.1ppm (n=8) |
|||
ColorChange |
Blue → Yellow |
|||
ReactionPrinciple |
SO2 +BaCl2 +H2O 仺 BaSO3 +2HCl |
|||
| Coefficient ofVariation | 10% (for2 to 10 ppm), 5% (for 10 to 30 ppm) | |||
| ShelfLife | 3Years | |||
| Correctionsfor temperature & humidity |
Unnecessary | |||
| Store the tubes at cool and darkplace. | ||||
| Possible coexisting substances and theirinterferences | (NOTE) |
Substance |
Concentration |
Interference |
Change color by itself |
Carbonmonoxide, |
- | No effect |
Nodiscoloration |
Carbondioxide |
100% |
Faintdemarcation |
100% CO2gives faint demarcation |
Nitrogendioxide |
1:1 |
Plus error |
Produces pale purplediscoloration |
Hydrogensulfide |
- | No effect |
Nodiscoloration |
| Othersubstance measurable with this detector tube |
Substance |
Correction Factor |
pump strokes |
Measuring range |
Thionylchloride |
0.72 |
2 |
1.44 to 21.6ppm |
| Calibration gasgeneration | Permeation tubemethod | |
| TLV-TWA | TLV-STEL | Explosive range |
| 2ppm | 5ppm | - |