Welcome Customer !

Membership

Help

Shanghai Yudong Electronic Technology Co., Ltd
Custom manufacturer

Main Products:

intelligent-mfg>Products

Shanghai Yudong Electronic Technology Co., Ltd

  • E-mail

    1375312008@qq.com

  • Phone

    18121126516(微信同号)13301881771

  • Address

    Building 10, 4th Floor, Block F, No. 1188 Lianhang Road, Minhang District, Shanghai

Contact Now

5La SO2 gas detection tube Sulphur dioxide

NegotiableUpdate on 06/06
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

5La SO2 gas detection tube Sulphur dioxide,Measuring Range 0.5 to 1 ppm 1 to 20 ppm 2 to 30 ppm 30 to 60 ppm, Portable gas detector, hydrogen alarm, all available on the China Environmental Emergency Monitoring Instrument Network. Emergency detection boxes for chemical accidents, nuclear and chemical emergencies, as well as emergency detection boxes for sudden incidents. Standard configuration includes detection tubes, water quality emergency detection boxes, emergency gas detection boxes for sudden incidents, and emergency detection boxes for sudden disasters in public places

Product Details

5La SO2 气体检测管 Sulphur dioxide

Sulphur dioxide No.5LA

SO2

5La SO2气体检测管 Sulphur dioxide

Performance

MeasuringRange

0.5 to 1 ppm

1 to 20 ppm

2 to 30 ppm

30 to 60 ppm

Number of PumpStrokes

8

4

2

1

CorrectionFactor

1/4

1/2

1

2

SamplingTime

1 minute per pumpstroke

DetectingLimit

0.1ppm (n=8)

ColorChange

Blue → Yellow

ReactionPrinciple

SO2 +BaCl2 +H2O 仺 BaSO3 +2HCl
HCl+Base 仺 Chloride

Coefficient ofVariation 10% (for2 to 10 ppm), 5% (for 10 to 30 ppm)
ShelfLife 3Years
Correctionsfor
temperature & humidity
Unnecessary
Store the tubes at cool and darkplace.

Possible coexisting substances and theirinterferences (NOTE)


Substance

Concentration

Interference

Change color by itself

Carbonmonoxide,
Nitric oxide

-

No effect

Nodiscoloration

Carbondioxide

100%

Faintdemarcation

100% CO2gives faint demarcation

Nitrogendioxide

1:1

Plus error

Produces pale purplediscoloration

Hydrogensulfide

-

No effect

Nodiscoloration



Othersubstance measurable with this detector tube


Substance

Correction Factor

pump strokes

Measuring range

Thionylchloride

0.72

2

1.44 to 21.6ppm



Calibration gasgeneration Permeation tubemethod



TLV-TWA TLV-STEL Explosive range
2ppm 5ppm -