What are the common problems with the adhesion tester
Date: 2025-08-29Read: 18
The common problems that may occur during the measurement process of the adhesion tester can be summarized into the following five categories, and their causes and solutions are as follows:1、 Unstable accuracy
cause:
-
Sensor aging or zero biasLong term use leads to a decrease in sensor sensitivity or uncorrected initial calibration deviation.
-
environmental interferenceTemperature and humidity fluctuations exceed the allowable range of the instrument, affecting material viscosity or sensor readings.
-
operational errorPoor fixture fixation and unstable testing speed (such as uneven manual operation) result in data fluctuations.
Solution:
- Regularly calibrate sensors and replace aging components.
- Strictly control the environmental temperature and humidity (such as using a constant temperature and humidity chamber).
- Adopting automated testing equipment to reduce human operational errors.
2、 Sensor malfunction
Common manifestations:
-
Abnormal readingThe displayed value is much higher or lower than the actual value, or there is no reading.
-
Calibration failureAfter calibration, accurate measurement cannot be restored.
cause:
- Sensors are affected by moisture, dust pollution, or physical damage (such as collisions).
- Poor sealing of the instrument leads to corrosion of internal components.
Solution:
- Check the sealing of the instrument and clean the surface stains of the sensor.
- Replace the damaged sensor and recalibrate it.
3、 Environmental interference
Key impacts:
-
temperatureHigh temperatures may soften materials, while low temperatures can reduce viscosity, both of which affect test results.
-
humidityHigh humidity environments may cause materials to absorb moisture and alter their adhesive properties.
-
test speedToo fast or too slow speed can cause measurement deviation in viscous force.
Solution:
- Follow standard testing conditions (such as ASTM D3330 specifying 25 ℃± 1 ℃, 50% RH ± 5%).
- Use a constant speed drive device to ensure stable testing speed.
4、 Data processing software issues
Common Faults:
-
Data read errorThe software is unable to recognize sensor signals or display garbled text.
-
record failureTest data not saved or saved in incorrect format.
cause:
- The software version is outdated and has poor compatibility.
- Storage device malfunction or file system damage.
Solution:
- Update the software to the latest version and check compatibility.
- Use reliable storage devices and regularly backup data.
5、 Improper operation and maintenance
Typical issues:
-
Loose fixtureDuring the testing process, the sample was displaced, resulting in invalid data.
-
Instrument leakageDamage to the insulation of the heating wire or motor can pose a safety hazard.
-
Insufficient cleaningResidual samples contaminate the sensor or testing area.
Solution:
- Check the tightness of the fixture before operation to ensure that the sample is securely fixed.
- Regularly inspect circuits and replace aging components.
- Clean the instrument promptly after testing to avoid sample residue.